start › cdms › k100 › run_diary › 20170406

2017/04/06 (Thursday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
0905 Tower/MC/Coldplate = 33.6/31.1/193 mK, Still = 1.3 K / 0.120 mbar, G1/G2 = 245/220 mbar, 1K Pot = 1.682 K / 28 mbar, Main bath = 36.0%. Boil off rate: faster than slow mode (predicted level: 39.8%)
0910 Fill LN2 cold trap bath
1347 Start LHe transfer @ main bath = 29.0%
1450 Finish LHe transfer @ main bath = 91.5%
New standard QET biases from most recent IbIs measurements:
channel QET bias prefactor QETbias*prefactor
DCRC1pA/DIB1pA/Side1pE 132.7 0.301 40
DCRC1pB/DIB1pB/Side1pB 145.3 0.275 40
DCRC1pC/DIB1pC/Side1pD 130.4 0.302 40
DCRC2pC/DIB4pC/Side1pC -152.8 0.234 40
DCRC2pD/DIB4pD/Side1pF -151.7 0.264 40
1655 Dial HV up to 20.0 V
1705 HV = 20.00 V, Vtest = 1.2 mV → Vdetector = 19.5 V
1708 Start sequence of 16, 1-hour data series of 1-sided 20V HV data with Frontend Gain = (20/20/20)/(-/4/4), OG = 4/2, QET biases = table above, with 1000 beginning randoms, no flashing, and a 15ADC bin threshold trigger on DCRC1 pA
1720 HV = 20.00 V, Vtest = 0.6 mV → Vdetector = 19.7 V
time Main bath recovery rate (cuft/min)
no Dewar on recovery
9:05 36.0% 0.228
9:30 35.4% 0.253
9:55 34.9% 0.213
11:05 33.0% 0.231
12:30 30.8% 0.168
13:10 29.6% 0.196
after transfer. Dewar on recovery
15:05 91.5% 0.447
15:20 91.5% 0.188
15:55 91.2% 0.234
16:45 90.8% 0.179
17:10 90.5% 0.159
17:44 90.0%
17:55 0.156
1739 Fill LN2 cold trap bath.
1747 Start taking 1 photo every 10 seconds of He recovery meter.