2017/03/14 (Pi day)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
0909 Tower/MC/Coldplate = 33.1/30.6/190 mK, Still = 1.2 K / 0.110 mbar, G1/G2 = 260/223 mbar, 1K Pot = 1.692 K / 28 mbar, Main bath = 70.1%. Boil off rate: close to slow mode (predicted level: 72.0%)
0912 Fill LN2 cold trap full.
0929 Ground detector and start 20 minute mini-bake.
1300 Fill HPGe LN2 dewar.
1635 Finish IbIs for DCRC2pC / DIB4pC / Side1pC
channel QET bias prefactor QETbias*prefactor Selected QETbias*prefactor New QET bias
DCRC1pA/DIB1pA/Side1pE 127 0.2129 27.04 40 130.2
DCRC1pB/DIB1pB/Side1pB 150 0.3514 52.71 40 143.4
DCRC1pC/DIB1pC/Side1pD 130 0.3014 39.18 40 130.4
DCRC2pC/DIB4pC/Side1pC -150 0.1611 24.16 40 -156.1
1755 Fill LN2 cold trap full.
1700 Start sequence of 12, 1-hour data series of 1-sided 0V HV data with Frontend Gain = (20/20/20)/(-/4/3), OG = 4/2, QET biases = table above, with 1000 beginning randoms, no flashing, and a 15ADC bin threshold trigger on DCRC1 pA