start › cdms › k100 › run_diary › 20170304

2017/03/04 (Saturday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
Overnight bake was stable at 480 mK, ended at 0518
1239 Tower/MC/Coldplate = 33.8/31.1/194mK, Still = 1.28 K / 0.110 mbar, G1/G2 = 249/222 mbar, 1K Pot = 1.691 K / 28 mbar, Main bath = 62.0%. Boil off rate: faster than slow mode (predicted level: 66.3%).
1243 Fill LN2 cold trap
1244 Tweak 1K pot needle valve
1346 1K pot pressure = 30 mbar
1349 Prebias detector to HV = 15 V
1405 Reduce HV to 10V
1407 Start sequence of 16, 1-hour data series of 1-sided 10V HV data with Frontend Gain = (20/20/5)/(6/4/3), OG = 8/4, with 1000 beginning and ending randoms, no flashing, and a 35ADC bin threshold trigger on DCRC1 pA
1409 HV = 10V, V_leak = -0.3 mV, V_crystal = 10.1 V
1419 Notice DIB2 PC railed.
1429 Recovered DIB2 PC.
1758 Pulses still look good