UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20161220

2016/12/20 (Tuesday)

Previous Day

Status: IVC open, 4K stage removed. Main bath vacuum jacket is evacuated. 1K Pot needle valve closed. Mixture has been pumped out of LN2 trap, traps are at room temperature, still need to be pumped out. Mixture is in dump, fridge evacuated.
G1 has drifted up to 55 mbar overnight, again not sure if this is from the outside air or maybe from the cold traps. Can rule out the latter by pumping the traps out today and seeing whether it goes up as much by tomorrow.
0845 Finish drilling second hole through the fridge port for the LHe trap. Takes a long time of constant drilling, maybe 20 minutes. Used #44 drill bit.
0900 Open Trap 1 to G1. G1 increases from 55.6 to 103.1, so 47.5 mbar in trap according to G1. Open to P2, after first pumping out P2 volume. P2 shows 45 mbar in trap. Pump out Trap 1 with 4He pump.
0910 Open Trap 2 to G1. I opened v12b first, and the G1 pressure immediately started to increase, indicating leak through the speedivalve, even though it was close pretty tightly. After opening the speedivalve G1 increases to 717.8 mbar, quite high. Open to P2, G1 falls to 295.8 mbar, P2 = 120 mbar (probably not very accurate for the Pirani gauge). Pump out Trap 2 with 4He pump.
1025 S101 shipping vessel pressure = 10.8 PSI, higher now that it's warm

Detector S101 DIB 1
date/who: 2016-12-20 NM MF
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 303 Ω
3 4 QET B 100 to 200 Ω 359 Ω
5 6 QET C 100 to 200 Ω 293 Ω
7 8 QET D 100 to 200 Ω 395 Ω
9 chassis LED1 Rforward 8 to 9 MΩ 560 kΩ
9 chassis LED1 Vforward 1.0 to 1.1 V 1.042 V
chassis 9 LED1 Rreverse <1 nS 200 kΩ
10 chassis LED2 Rforward 8 to 9 MΩ 767 kΩ
10 chassis LED2 Vforward 1.0 to 1.1 V 1.045 V
chassis 10 LED2 Rreverse <1 nS <0.3 nS
11 12 Qo-Qi n/a
12 11 Qi-Qo n/a
1-8 chassis <1 nS <0.1 nS
Detector S101 DIB 3
date/who: 2016-12-20 NM MF
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 13 MΩ
3 4 QET B 100 to 200 Ω open
5 6 QET C 100 to 200 Ω open
7 8 QET D 100 to 200 Ω open
9 chassis LED1 Rforward 8 to 9 MΩ open
9 chassis LED1 Vforward 1.0 to 1.1 V open
chassis 9 LED1 Rreverse <1 nS open
10 chassis LED2 Rforward 8 to 9 MΩ open
10 chassis LED2 Vforward 1.0 to 1.1 V open
chassis 10 LED2 Rreverse <1 nS open
11 12 Qo-Qi n/a
12 11 Qi-Qo n/a
1-8 chassis <1 nS <0.11 nS
Additional check of continuity between QETA and QETC
1 5 QET A-C 3.97 kΩ
1 6 QET A-C open
2 5 QET A-C 702 kΩ
2 6 QET A-C open
Detector S101 DIB 4
date/who: 2016-12-20 NM MF
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 349 Ω
3 4 QET B 100 to 200 Ω 420 Ω
5 6 QET C 100 to 200 Ω 347 Ω
7 8 QET D 100 to 200 Ω 378 Ω
9 chassis LED1 Rforward 8 to 9 MΩ open
9 chassis LED1 Vforward 1.0 to 1.1 V open
chassis 9 LED1 Rreverse <1 nS open
10 chassis LED2 Rforward 8 to 9 MΩ open
10 chassis LED2 Vforward 1.0 to 1.1 V open
chassis 10 LED2 Rreverse <1 nS open
11 12 Qo-Qi n/a
12 11 Qi-Qo n/a
1-8 chassis <1 nS <0.1 nS
Side 1 DIB 1 Side 1 DIB 3 Side 1 DIB 4
Side 2 DIB 1 Side 2 DIB 3 Side 2 DIB 4
1110 Put S101 in purge cabinet
1340 Begin removing 100 mK stage

In this run the cold plate temperature seemed higher than it should be. I suspected a touch between the 100mK stage and a warmer stage. The 4K stripline “buckle” seemed to be free of the 100mK can. The copper tape around the 600mK rod hole in the 100mK shield did not seem to be touching. The other possible culprit is between the 100mK shields or heat sinks and the 600mK face of the tower. Note that the 100mK section of the tower is almost completely internal, with only the ears sticking out - above the ears is the MC stage and below is the 600mK stage. So if the 100mK heat sinks or shields touch the tower face that's a contact between the 100mK and 600mK stages. The pictures above show no obvious touch but clearly things get pretty close.

1420 Finish removing tower, place tower in clean room
1515 Put tower in N2 purge cabinet

Next Day