2016/11/17 (Thursday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters → cables1/2/3 → DCRC55(1)/33(2)/50(3)
0810 Tower/MC/Coldplate = 32.9/35.7/210 mK, Still = 1.1 K / 110 μbar, G1/G2 = 185/185 mbar, 1K Pot = 1.69 K / 26 mbar. MB = 63.3%.
Boiloff rate: Very close to slow-mode, still a little bit faster
0815 Fill LN2 coldtrap bath. LN2 Dewar is now empty.
1600 The DCRC driver was crashing whenever DCRC2 was enabled. This problem was fixed by not requiring DCRC2 (aka DCRC33) to read its calibration file. That calibration is not needed for this run but in the future we might want it again. In the ODB the key to do this is / Equipment / Tower01 / Settings / DCRC2 / General / UseCalibrationRegisters, set to “n”
1654 Fill LN2 coldtrap full.
1742 Test series for background data with NaI trigger. Same conditions as two days ago and trying for the same QET biasing scheme.
1749 Start sequence of 15 one-hour data series of 0V data with Frontend Gain = 5/2/5, OG = 8/3/8, with 2000 beginning randoms (because of increased pileup from higher source rate), with no flashing in between series, triggered with NaI coincidence setup (low threshold 92mV high threshold 124mV and 1100V on NaI detector) using DCRCX (DCRC4) PhA.
1750 MB = 50.1%