UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20161110

2016/11/10 (Thursday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters → cables1/2/3 → DCRC55(1)/33(2)/50(3)
Last night's data looked OK in the first series at 16:06, but by the second series at 17:06 the noise in DCRC3pD, which we were triggering on, must have increased, so that most of the pulses are noise triggers. Then in the middle of the run at 23:07, DCRC1pD railed. (That's an inner ring channel on Side 2 so probably not the most critical one for analysis.)
0900 Fill LN2 cold trap bath.
0930 Tower/MC/Coldplate = 32.8/35.9/232 mK, Still = 1.37 K / 110 μbar, G1/G2 = 175/165 mbar, 1K Pot = 1.70 K / 28 mbar. MB = 21.6%.
Boiloff rate:
0951 Start LHe transfer @ 20.7%
1045 Finish LHe transfer @ 91.5%
1140 Unrail DCRC1pD using “Auto Zero once” on the SQUID Tuning gui. Then set its baseline back down to -0.8 V
1145 Check event rate, count 33 events in 1 second = 33 Hz. Down from 46 Hz yesterday, 110 Hz the day before.
1200 DIB2pC seemed to have high 10kHz oscillations so tried adjusting its QET bias downward. Sometime during this process, all other channels railed! After restoring them they had no pulses, probably superconducting. Annoying, I don't know what happened. Possibly an operator error caused a glitch. For example I tried Auto Zero on DCRC2, which I know does not work, apparently because it's RevC not RevC.1. Maybe this does something weird.
1205 Release pressure from main bath
1210 On DCRCpA tried “snapping” QET bias to 800, then down to 200 - at 200 saw pulses. Then to its previous value of 179. But this put it into superconducting. So maybe this is one of those situations where suddenly a previous parasitic power source disappeared, so that now all the channels will require a higher bias.
1355 DIB1pA seemed to require a QET bias 3 μA higher than previous. Try applying 3 μA higher QET biases to all DCRC1 channels - seems to have work, all have good pulse shapes and reasonably low 10kHz oscillations
1410 On DCRC2 & DCRC3 the previous QET biases seem to work still.
DCRC Phonon
channel
SQUID
lockpoint
SQUID
bias
QET
bias
Updated
QET
bias
1 0 -3.1 -76.5 179 182
1 1 -3.72 -81.2 182 185
1 2 -3.5 -70 182 185
1 3 -1.8 -82.6 187 190
2 0 1.67 50 0
2 1 2 64.9 -174
2 2 3.07 60.3 -173
2 3 1.5 60.6 0
3 0 -4.68 -84.8 0
3 1 -3.06 -86.7 207
3 2 -0.9 -48 165
3 3 -4.7 -51.3 202
1415 Experiment with trigger settings on DCRC3pD. Threshold = 70 seems to be mostly above noise, only occasionally does it seem to trigger on noise. Possibly a little on the high end but we'll try it.
1425 Adjust driver offsets to get all DC baselines around -0.8 V
1428 Start sequence of 15 one-hour data series of 0V data with Frontend Gain = 5/2/5, OG = 8/3/8, with 1000 beginning randoms, with no flashing in between series, and a 70ADC bin threshold trigger on DCRC3 pD.
1500 Data looks OK so far, except DIB1pC pulses have 1 kHz “wiggles”, usually manifesting as a second peak but sometimes multiple peaks are seen. Is this phase separation?
DIB1pC (Side1pD) pulse example with wiggles
1655 Fill LN2 cold trap bath
1715 Main bath = 85.8%

Next Day