2016/10/24 (Monday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters → cables → DCRC55(1)/33(2)/50(3)
0830 Return Cf source to safe.
0832 Tower/MC/Coldplate = 34.0/36.2/206 mK, Still = 1.07 K / 85 μbar, G1/G2 = 121/100 mbar, 1K Pot = 1.55 K / 4.5 mbar. MB = 36.2%. 1K pot looks to have been stable over the last 12 hours.
Boiloff rate is good, slow-mode
0838 Fill LN2 cold trap full.
0846 Tweak 1K pot needle valve.
0905 Unrail all channels.
0910 Measure event rate by counting pulses above noise in DIB4pD. 41 pulses counted in 0.5 seconds (ten 50 ms traces), rate approximately 81 Hz.
At this rate, about 20% of random traces would be pulse-free with 20 ms traces. About 1.5% would be pulse-free with 50 ms traces.
0920 MIDAS DAQ webpage has crashed. Attempting to restart it.
0925 Restarted DAQ using instructions Starting the MIDAS DAQ Website
0925 I think a data series was started by accident after the DAQ restarted.
0930 Check channels again, all look OK
0930 Start sequence of one-hour data series of 0V data with Frontend Gain = 5/2/5, OG = 8/3/8, with 1000 beginning and ending randoms, with no flashing in between series, and a 40ADC bin threshold trigger on DIB1 pB. (It's the 15-hour sequence but I plan to stop it early.)
1255 Commence LHe transfer (29.1%).
1257 Stop data taking, a little late. The last few minutes of the 12:30 series will probably be noisy.
1333 End LHe transfer (91.6%).
1525 Check phonon channels, all look OK.
1630 Start data series of 0V data with Frontend Gain = 5/2/5, OG = 8/3/8, with 36000 beginning randoms, with no flashing, with all trace lengths set to 20 ms. Intent is to take long-trace randoms only for noise studies. The randoms rate is 20 Hz for 30 minutes of randoms so I'll stop the data manually after 30 minutes.
Note: there are about 350 events per dump in this data. Dumps are 1 minute so there are actually 1200 randoms generated per dump. So only about 30% of the randoms generated are produced. Or in other words the read rate is only about 6 Hz. That's for 18 20-ms traces per event.
1635 Let pressure out of main bath. Forgot to do it earlier.
1703 Stop data.
1715 Fill Ln2 cold trap full.
1715 1K Pot is down to 1.55 K / 5.5 mbar. Tweak needle valve open a few more degrees.
1740 Put Cf source next to fridge for overnight activation.