2016/10/01 (Saturday)

Previous Day

DCRC1 and DCRC3 powered on and connected through HV setup EBoxDIB 2/3 → 3.2 MHz filters → Cables → remapping board I/remapping board II → HVside2board “DIB1”/“DIB2” → DCRC55(1)/50(3). Remapping boards have all possible lines switched off. .
Voltage sweep running until ~0640.
0638 Commence 2 hour UV bake. Fridge should be cold again before morning data taking. (<100 mK).
1030 Fill LN2 coldtrap full.
1050 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1111 Cease data run, ground detector. Flash with UV LEDs for 30s, and cool down for 10min.
1125 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1147 Cease data run, ground detector. Flash with UV LEDs for 30s, and cool down for 10min.
1201 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1222 noticed base temperature is up a bit, to 84.2mK
1224 Cease data run, ground detector. Flash with UV LEDs for 30s, and cool down for 10min.
1237 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1257 Cease data run, ground detector. Flash with UV LEDs for 30s, and cool down for 10min.
1310 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1333 Cease data run, ground detector. Flash with UV LEDs for 30s, and cool down for 10min.
1350 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1410 Cease data run, ground detector. Flash with UV LEDs for 30s, and cool down for 10min.
1430 Fill HPGe Dewar
1434 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ 0V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1458 Cease data run, ground detector. Flash with UV LEDs for 1min, and cool down for 10min. Tower temp peaks at 256mK.
1516 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ 0V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1539 Cease data run, ground detector. Flash with UV LEDs for 30s, and cool down for 10min.
1553 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +24V HV/ -12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1555 Tower/MC/Coldplate = 85/55/??(2.28 kΩ) mK, Still = 1.11 K / 90 μbar, G1/G2 = 208/190 mbar, 1K Pot = 1.58 K / 6.5 mbar. MB = 64.6%
1614 Cease data run, ground detector. Without flashing, start new series: With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with -24V HV/ +12V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1642 Cease data run, ground detector. Switch to IR LEDs by changing diverter board switches and modifying current values in flashandflash.py. Flash with IR LEDs for 30s, and cool down for 10min. Tower temp peaks at 287mK.
1708 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ 0V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1730 Cease data run, ground detector. Flash with IR LEDs for 30s, and cool down for 10min.
1735 Fill LN2 cold trap full.
1752 With 33kΩ resistor on DIB1 QOFB line, manually start 20 minute series of data with +12V HV/ 0V DIB1 with 1000 starting/1000 ending randoms with 220 ADC Bin trigger threshold on DIB1 Qo. no source
1812 Cease data run, ground detector.
1816 Begin sequence of overnight bakes and dcrc voltage scan using “Alternating_flashandsweep_0to12_and_12to0_evenV_IR.msl”. I believe it looks like the following: 2hr IR bake; 1 hr cooldown; 15 minute series with 30s/10min IR flash/cooldown at HV=0V, DCRC=0/0/-2/2/-4/4/-6/6/-8/8/-10/10/-12/12 V ; 2hr IR bake; 1 hr cooldown; 15 minute series with 30s/10min IR flash/cooldown at HV=0V, DCRC=-12/12/-10/10/-8/8/-6/6/-4/4/-2/2/0/0 V.