2016/09/23 (Friday)
| DCRC1 and DCRC3 powered on and connected through HV setup EBoxDIB 2/3 → 3.2 MHz filters → Cables → remapping board I/remapping board II → HVside2board “DIB1”/“DIB2” → DCRC55(1)/50(3). Remapping boards have all possible lines switched off. . | |
| Baked using UV LEDs for 8 hours last night. Tower was at 0.40 K until 06:53 then fell to base over about 40 minutes. | |
| 0835 | Tower/MC/Coldplate = 93/57/??(2.1 kΩ) mK, Still = 1.12 K / 95 μbar, G1/G2 = 175/160 mbar, 1K Pot = 1.56 K / 5.5 mbar. MB = 45.5% As yesterday the thermometers connected to “FRIDGE OUT 1”/“SENSOR 1” cable to the thermometer breakout box seem to be reading wrong. Coldplate is off scale, Still is higher than usual, MC Sensor = 79 mK Boiloff rate: good 1K Pot has been stable. |
| 0825 | Fill LN2 cold trap bath. |
| 1009 | Start data, but forgot to put out Ba source. Stop data, ground detector, and put out Ba source. |
| 1017 | Manually start 15min background data series at HV = 8V, DCRC1/Readout= -8V, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo. |
| 1032 | Cease data tests. Ground detector, flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1045 | Manually start 15min background data series at HV = -8V, DCRC1/Readout= 8V, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo. |
| 1101 | Cease data tests. Ground detector, flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1115 | Manually start 15min background data series at HV = 7V, DCRC1/Readout= -7V, with 1000 initial randoms, 1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo. |
| 1133 | Cease data tests. Ground detector, flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1150 | Return Ba source to safe |
| 1200 | Do a rate check with no external source: apply 10 V on DCRC1qO on the DCRC, expand pulseviewer window to 50 ms, take some random traces and count how many visible pulses there are. In 400 mS 56 pulses seen, so rate is 140 Hz. |
| 1211 | Flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1230 | Move the 1 Ci PuBe source into the lab. It is in its paraffin-filled transport barrel on a stool behind the fridge on the west wall (by the LN2 cold traps). One of the movable shield walls has been moved into the west position between the PuBe source and the fridge. According to Andy Astleford the background is low enough to work at all of our work stations. |
| 1240 | Remeasure event rate. Counted 42 pulses in 400 ms, even less than before. I may not been counting as many small pulses as before, so maybe there's a bias: could be about 10 more pulses. So the rate with the PuBe source is no bigger than the rate without the source. |
| 1255 | Flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1309 | Manually start 15min background data series at HV = -7V, DCRC1/Readout= 7V, with |
| 1325 | Cease data tests. Ground detector, flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1339 | Manually start 15min background data series at HV = 6V, DCRC1/Readout= -6V, with |
| 1346 | Brief base temp excursion from 90mK to 110mK. |
| 1356 | Cease data tests. Ground detector, flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1411 | Manually start 15min background data series at HV = -6V, DCRC1/Readout= 6V, with |
| 1427 | Cease data tests. Ground detector, flash with UV LEDs for 30s, then cool down for 10 minutes. |
| 1445 | Manually start 15min background data series at HV = 4V, DCRC1/Readout= -4V, with |
| 1505 | Crap, trace length was still long since looking at the event rate at 1240. So the last four series have 50 ms traces and also were set to 100000 initial randoms, so it's all randoms. |
| 1526 | Commence LHe transfer (34.1%). |
| 1614 | End LHe transfer (92.8%). |
| 1625 | Release pressure from main bath. |
| 1650 | Release pressure from main bath. |
| 1700 | Remove resistor from DIB1 QOFB line and commence automated eight series of 30 sec flash with UV LEDs/10 minute cooldown/20 minutes data with 0V HV/ 0V DIB1 and 1000 starting/1000 ending randoms with 150 ADC Bin trigger threshold on DIB1 Qo. |
| 1704 | Fill LN2 cold trap full. |
| 1715 | Release pressure from main bath. |
| 1900 | Na-22 source placed 2-3 feet from center of main bath, on the floor. |
| 2114 | Commence 12 hour UV bake. |