2016/09/18 (Sunday)

Previous Day

Status: LHe in bath, mixture circulating, DCRC1 and DCRC3 powered on and connected through HV setup EBoxDIB 2/3 → 3.2 MHz filters → Cables → remapping board I/remapping board II → HVside2board “DIB1”/“DIB2” → DCRC55(1)/50(3). Remapping boards have all possible lines switched off. .
Baked using UV LEDs for 12 hours last night
1338 MC(Cernox)/Tower/1K Pot = 60 mK/90 mK/1.44 K,P2 = 2.0 mbar G1/G2 = 115/1117 mbar, MB = 48.5%
Fill LN2 cold trap full. Tweak 1K Pot needle valve.
1342 Noise still seen on DCRC1 Qo, with no voltage applied. Applying +4V to Qouter results in pulses similar to those posted yesterday which feature the damped harmonic effect.
1358 Return to DAQ after 15 minutes to find pulses no longer triggering. Reducing the trigger value to 3000 shows very fuzzy noise, and no pulses. Switching the Qo bias to -4V makes pulses reappear, though they are fuzzier at times. Perhaps we can only take 10-15 minutes of data at a time?
1407 Pulses are starting to breakdown. It appears that pulses were clean for 10 minutes.
1458 After some fiddling with the switchboard but not wanting to solder permanent changes on to it, we have given up. Power up the boards and start a 12 hour bake with IR.
1508 Tweak needle valve one last time.
1510 G1/G2 = 142/143 mbar, P1=0. 14 mbar, 1K Pot = 2.25 K / 1 bar (off charts) and falling, Tower temperature at 455 mK and still slowly rising due to bake. DCRCs are still powered on in HV setup. He level at 46.4%.
Boiloff rate bad. Yesterday pressure was only released one time, 7 minutes after a 90.5% fill.