2016/08/29 (Monday)
| Setup: DIB1/2/3 → 3.2 MHz filters → nothing/disconnector board/remapping board → HVside2board → DCRC55(1)/50(3). Disconnector board has all charge lines disconnected. | |
| DCRCs powered overnight. Not pumping on IVC.. Stopped circulation, and 1K Pot to allow warm up. | |
| 0915 | Tower= 4.6 K, Still = 24 μbar, G1/G2 = 9.0/548 mbar (V9 closed), 1K Pot = 4.2 K / 0.5 mbar, MB = 1.2%. IVC=2.2 μbar. |
| 0950 | Lower Dewar. |
| 0954 | With tower temperature of 40K after lowering dewar, IVC=0.14 mbar. Turn on leak checker. |
| 1030 | Perform electrical check with tower at 80K. All measurements as before except those mentioned below. |
| DIB1: QIbias→QObias=10nS, QIbias→QETret<1nS QIfb→Drain/Source <1nS, QIfb→chassis=14Ω QETrets→chassis~3nS QETB→Ret = 200Ω |
|
| DIB2: QETB→ret=open, 70Ω, open, 71Ω QIfb→Drain/Source <1nS, QIfb→chassis <1nS QOfb→Drain/Source=20nS, QOfb→chassis<1nS FET Tsense+→FET heater- = open, QET rets→chassis ~3nS |
|
| DIB3: Squid Bias D/ret unstable (2-20 kΩ) QETs/QETrets→chassis ~3nS |
|
| 1044 | Filling LN2 cold trap to keep the gunk frozen. |
| 1054 | 8.5mbar was in P1 due to fridge warm up, pumping it to dump. |
| 1155 | Tower temp = 195K. IVC pressure = 1.9mbar. Commence pumping on IVC with scroll pump. |
| 1205 | Switch to turbopumping IVC. |
| G2: 775mbar. Commence pumping whatever may be left in LN2 cold traps to dump (both cold trap 1 & 2) through V12A/B, both cold trap speedivalves, and V5. | |
| 1335 | Close V12A/B and cold trap speedivalves, V5, and begin closing V6 in preparation for throughput test. G1: ~1mbar, G2: 780mbar, P1: 6ubar. |
| Calibrate Leak Checker and open it to the IVC. Isolate Turbopump, so we're only pumping on the IVC with the leak checker. | |
| 1340 | Background leak rate stabilizes at 1.1E-6 mbarL/s. Restart pumping on IVC with turbopump. |
| 1436 | G2: 783mbar in dump. Isolate dump. Pump out lines leading from P2 to Cold Traps 1 & 2 (V5A, 2A, V7). Once pumped down, close manual valve on He4 pump and power off. Open V11A/B to open Cold Traps to P2. Remove cold traps from LN2 bath and allow to warm up. |
| 1510 | With IVC pressure down to 8.6e-5 mbar, attempt to switch to leak checker to obtain background rate and start leak check. Background rate was at ~1e-6 mbarL/s which is TOO DAMN HIGH. Switch back to turbopump, and see that IVC pressure has risen to 8.6E-4 mbar. Not sure if this is due to switching to the leak checker and back, or due to temperatures rising. Let's leave this alone a while longer. |
| 1520 | Perform an electrical check with tower at 230K. All measurements as before except those mentioned below. |
| DIB1: QIbias→QObias=13nS, QIbias→QET ret=25nS QIfb→Drain/Source <1nS, QIfb→chassis=2.5Ω QET ret→chassis~6nS SquidC/SquidC ret→chassis~6nS QETB→ret = 5MΩ or open Last QETB ret→chassis 2Ω (short) QiBias→chassis=6 nS and QiBias/QoBias→QET ret=25 nS |
|
| DIB2: QETB→ret=open, 89Ω, 103Ω, 90Ω QIbias/Qobias→ret=25 nS (instead of <1 nS) QIfb→Drain/Source=23/23nS, QIfb→chassis <1nS QOfb→Drain/Source=25nS, QOfb→chassis<1nS FET Tsense+→FET heater- = 0.75V (normal), QET rets→chassis ~5nS SquidA/SquidA ret→chassis=5 nS |
|
| DIB3: Squid Bias D→ret (now stable) 3.1kΩ QETs/QETrets→chassis ~6nS QIfb/QOfb→chassis <1nS Bfb/Bfb ret→chassis~6nS |
|
| 1538 | P2 reads 45mbar. Turn on He4 pump and open manual valve to begin pumping on warm cold traps. |
| 1605 | With tower temperature at 234 K and background leak rate of 3.4e-7 mbarL/s (IVC pressure down to 5.4e-4 mbar), conduct spray test on IVC cans and upper flanges. No sign of any leak, with leak rate slowly decreasing over time to 3.2e-7 mbarL/s. |
| 1627 | Cease pumping on LN2 cold traps. Close all valves on IGH, and return traps to LN2 bath. Fill LN2 cold trap full. |
| 1639 | Turn off mixing chamber heater and still heater in preparation for throughput leak check. Background leak rate at 3.5e-7 mbarL/s. |
| 1645 | Conduct throughput leak check through cold trap2, closing gate valve to force mixture to linger in the fridge lines. No leak/spike in leak rate was detected. Open gate valve and V3 to fully flush the mix through the fridge lines, still no indication of a leak. Commence pumping mixture back to dump through both sides of the fridge. |
| 1658 | Add 10 mbar N2 exchange gas to IVC to warm up overnight. Tower temperature currently at 237 K. |
| 1744 | Tower temp = 251K. G2 at 786mbar. Close V9 and turn off He3 pump. |