UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20160804

2016/08/04 (Thursday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55(1)/33(2)/50(3), Tower has been cooled below Al transition with mu-metal shield at standard position
DCRCs unpowered overnight
0705 Tower/MC/Coldplate = 38.9/34(?)/143 mK, Still = 1.16 / 0.050 mbar, G1/G2 = 150/145 mbar, 1K Pot = 1.40 K / 1.4 mbar, MB = 22.4%
0710 Tweak open 1K Pot needle valve a few degrees
0715 Fill LN2 cold trap bath
0718 Begin LHe transfer
0757 End LHe transfer @ 77.0%
0800 Tower = 36.4 mK, Still = 0.065 mbar, G1/G2 = 180/175 mbar, 1K Pot = 1.52 K / 5.5 mbar. Tweak 1K Pot needle valve open a little more
0807 Release pressure from MB, 10 min after transfer
0833 Release pressure from MB, 36 min after transfer
0835 Tower temperature spikes for no reason, to 45 mK. 1K Pot back down to 5.5 mbar, circulation and still pressures falling - tweak 1K Pot needle valve open again.
0840 Tower temperature falling again
notes on packaging detector stack from iT4 for transport to SD
1110 Tower temp = 38.5mK. P2: 3.4 mbar. Tweak open 1K Pot needle valve.
1130 Tower temp has risen to 42mK, though the 1K Pot seems to still be filled. P1: 90ubar. Turn on Still Heater to 1.0mW.
1133 Tower temp = 39.8mK. P1: 100ubar. Increase Still Heater to 2.0mW.
1136 Tower temp = 39.6mK. P1: 120ubar.
1138 Tower temp = 39.2mK. P1: 120ubar.
1143 Tower temp = 38.3mK. P1: 120ubar.
1145 Perform mini-electrical check on DIB1 to compare w/ the charge channels measured on DIB2 yesterday. Both Qi and Qo Feedback to Drain/Source measure ~18nS like DIB2 Qi did. So DIB2 Qo was the problem channel, it would seem.
1148 Tower temp stabilizing at 38.2mK. P1: 120ubar.
1158 Tower temp = 37.7mK. P1: 120ubar.
1245 Went into clean room and hooked up N2 supply line to Tower transport canister to flush with N2 gas. Filled with N2 gas to 15psi then opened to let pressure decrease to just above 1atm. Did this 15 times, which should be more than enough. On last purge, let the pressure fall to 8psi, then close valve to isolate canister.
Tower temp = 36.5mK and holding relatively stable.
1340 Place piece of tape on pressure gauge of Tower transport canister to mark the 8psi I filled it with. This will let us keep an eye on whether it's slowly losing pressure.
Tower temp = 35.7mK and steady. P1 = 110ubar, G1/G2 = 235/223 mbar, 1K Pot = 1.56K / 7.0mbar. May need to tweak open 1K Pot soon.
1430 Tower temp spikes from 36mK to 44mK, then begins decreasing.
1435 Tower temp back down to 37.0mK and still coming down. P2 down at 6.0mbar. Tweak open 1K Pot needle valve.
1455 Tower temp = 35.5mK. P1 = 120ubar, G1/G2 = 251/238 mbar, 1K Pot = 1.64K / 22mbar.
1457 Power on DCRC1. Tower temp spikes to 40.3mK then begins dropping.
1535 Tower temp = 37.4mK after dropping to ~36mK after the DCRC had been powered. Now it's slowly rising. P1: 120ubar, so that seems fine. The 1K Pot = 1.58K / 9mbar, so that might need adjustment. The DCRC1 phonon channels look good, but we need to zap. Let's see how that affects temperature.
1545 Tower temp = 37.8mK and still cooling down. Though P2 has dropped to 8mbar, and P1 is at 110ubar. Tweak open 1K Pot a little more. May need to raise Still Heater another mW.
1813 P2: 9mbar. Tweak open 1K Pot needle valve. Tower temp 37.4mK.
1820 Fill LN2 cold trap full.

Next Day