start › cdms › k100 › run_diary › 20160705

2016/07/05 (Tuesday)

Previous Day

Status Warming after finding a blockage. MB full of He gas, IVC has He transfer gas
0825 MC(Cernox) = 182.6 K, Sorb = >250 K, IVC = 2.5 mbar
0830 Check pressure in LHe cold trap by opening V1 and V3. P1 goes up slightly, from 0.017 to 0.020 mbar
0835 Open still gate valve and condenser speedivalve. No increase seen on P1.
0845 Fill LN2 cold trap bath
0920 Try a forward throughput test through V9, V13A, V12A, V1, still gate valve, then open condenser speedivalve. No increase on P1. No throughput.
0930 Pump mixture back to dump. Dump (G2) = 794.6 mbar.
0935 Let one hose-length of He gas into MB vacuum jacket to speed up warming.
1000 MC starts warming more quickly.
1100 MC(Cernox) = 192.8 K, IVC = 3.0 mbar
1245 Checked to see if there was oil in the still pumping line. First fill pumping line with clean N2 gas, through vent & V7 & V8 with V2 also open to monitor pressure, fill to 1000 mbar. Then disconnect still pumping line near 3He pump, just above V6. Use a wipe to check for oil - it's very dry and clean, no oil. Then disconnect near the gate valve, also very clean. Reconnect both o-ring joints then pump out with scroll pump through V5A & V2A.
1305 Stop pumping out still pumping line, got P1 down to 0.5 mbar, G1 down to 1.2 mbar/ P2 down to 0.019 mbar (P2 falls to 0.012 mbar after closing valves)
1320 Pump GHS contents (G1 and P1 volumes) to dump using 3He pump. Dump (G2) = 790.8 mbar
1330 MC(Cernox) = 215 K, IVC = 3.4 mbar
1330 Open LHe trap to P1 through V1, V3. P1 goes up from 0.014 to 0.032 mbar.
1335 Open condenser to P1 through speedivalve, P1 goes up from 0.022 to 0.034 mbar.
1340 Open still gate valve to P1. P1 goes up slowly from 0.024 to 0.038 mbar.
1345 Pump fridge to dump. Dump (G2) = 793.6 mbar.
1350 MC(Cernox) = 216.6 K.
1350 Try a forward throughput test through LN2 cold trap 1. P1 doesn't budge, no throughput.
1355 Pump fridge to dump
1425 Dump (G2) = 796.3 mbar.
1455 Commence cryocleaning through Cold Trap 2. V6 open to 12.5%.
1510 Close V13B to check flow. No flow. Increase V6 to 13.0% before flow is restored.
1620 Close V13B to check flow. Very slight flow that may be too slow. Increase V6 to 13.1%.
1645 Flow still present. Cease cryocleaning and pump to dump.
1730 Fill LN2 cold trap full. Cease pumping to dump. Isolate dump and turn off He3 pump.
1740 Perform forward throughput test. Before opening Gate valve, P1 read 16ubar. After opening Gate Valve and allowing mixture to flow through fridge into still, P1 begins climbing.
1745 P1 has climbed to 3.8mbar and is still increasing. Pump mixture to dump.

What was the temperature? Not recorded. Extrapolating from 13:30 and 13:50 it was probably about 235 K at the MC.

Attempting to pump mix to dump reveals that He3 pump is not running despite showing that it's on at the IGH. The electrical box in the pump alley shows DC Fail. Resetting and unplugging the various pumps does not help. Every time after turning the box on and then attempting to press the power button, the DC Fail light turns on. Will leave mixture chilling out in the fridge, cold traps, and dumps so that no pressures become too high overnight.

Next Day