start › cdms › k100 › run_diary › 20160629

2016/06/29 (Wednesday)

Previous Day

Status Tower installed. Two stripline breaks. One on Stripline 20 / DIB3, which means we have to try to run with its rerouting board next to DIB2's rerouting board.
0900 Check depth or SQUET-end rerouting board sockets. They are actually just the right size for the SQUET pins. I'm not sure how the stripline plug winds up seeming so much thinner.
1015 Finish Stripline 19 rerouting boards
1020 For the Stripline 20 D50-side rerouting board the previous jumper wire had been removed. Looking at the SQUET-end board, 21 is jumped to 11 with the 11 socket removed. That's qO Bias rerouted through QET A Return #3. So to restore the D50 side:
  1. for D50-end rerouting card:
    1. jump Pin 4 (QET A Return #3) on the male side (stripline side) to Pin 7 (qO Bias) on the female side (ebox side)
    2. open switches (to break connections) corresponding to Pin 4 and Pin 7
1100 Finish Stripline 20 rerouting boards
1300 Check the stripline continuity of the two broken traces after attaching rerouting boards to stripline ends. Both look good.
1310 Plug in DIB1 SQUET. Unfortunately SQUID B is open yet again. Did a pin-by-pin check of all Stripline 19 traces from BOB to stripline connector, checked all 50 with rerouting board in place. Find that SQUID B bias line is bad (pin 27 on SQUET side, pin 32 on D50 side). Trace resistance unstable when stripline is manipulated.

Plan for rerouting Stripline 19: Bad wire = SQUID B Bias. Reroute with good spare wire = QET D Return #1

  1. for SQUET-end rerouting card:
    1. add jumper from Pin 27 (SQUID B Bias) to Pin 40 (QET D Return #1)
    2. break off Socket 40
  2. for D50-end rerouting card:
    1. jump Pin 48 (QET D Return #1) on the male side (stripline side) to Pin 32 (SQUID B Bias) on the female side (ebox side)
    2. open switches (to break connections) corresponding to Pin 48 (switch U09-1) and Pin 32 (switch U09-2)
1400 Plugged in the DIB3 SQUET, did an electrical checkout. SQUID B reads very high resistance, 600 kΩ. Check stripline continuity - all looks ok. Check SQUID B on SQUET pins - all looks OK. Try plugging it in again.
1430 DIB3 plugged in, now SQUID B looks fine.
1600 Finish Stripline 19 rerouting boards
1630 Plug in DIB1 SQUET. Feedback B/RET is open. Find that Feedback coil B line is bad (pin 14 on SQUET side). Trace resistance unstable while moving stripline.

Plan for rerouting Stripline 19: Bad wire = Feedback Coil B. Reroute with good spare wire = QET C Return #2

  1. for SQUET-end rerouting card:
    1. add jumper from Pin 14 (Feedback Coil B) to Pin 7 (QET C Return #2)
    2. break off Socket 7
  2. for D50-end rerouting card:
    1. jump Pin 9 (QET C Return #2) on the male side (stripline side) to Pin 16 (Feedback Coil B) on the female side (ebox side)
    2. open switches (to break connections) corresponding to Pin 9 (switch U05-2) and Pin 16 (switch U10-2)
1725 Finish Stripline 19 rerouting boards

Next Day