| 1500 | Zap SQUIDs for the heck of it. |
| 1515 | Commence LHe transfer (30.6%). |
| 1515 | Reestablish timing link, and check charge channels. Triangular noise still present. |
| 1540 | Try biasing charge channels to 4V with the DCRCs. Pulses present above the noise. This is with nothing plugged into the HV coax, so this is the lowest noise state. |
| | With HV input plugged into the HV supply (with the supply off) you can't see the pulses above the large noise. |
| 1550 | Cease LHe transfer (93.1%). |
| 1650 | Let pressure out of main bath. |
| 1655 | Before switching back to IZIP mode, disable DCRC1. Unpower DCRC1. Noise still present on DIB2. Unplug cable to EBox from the back of the HV board → The noise goes away on DIB2. Plugging this cable back into the back of the HV board causes the noise to reappear. |
| | Lifting the HV Box and unpowered DCRC1 up off the top of the Break Out Box causes the noise to go away, but when we try to figure out if it's the DCRC connection to the box or if it's the HV Box touching the Break Out Box, we can't get the noise to go away again. |
| | Removing the 3.2MHz filter has no effect. Swapping the order of the gender swapper and the 3.2MHz filter has no effect. |
| | Switching the cable from the HV box to the EBox with another shielded cable has no effect. |
| | Touching the shell of any one of the DSUB connectors on the HV Box to the metal chassis of the HV Box causes the triangular noise to go away. Boom! |
| 1725 | Repower DCRC1 attached in the IZIP setup: EBox → Cable → Gender Swapper → Filter → DCRC. Pulses look good. |
| 1735 | Fill LN2 cold trap full. |
| | 1K Pot has been pretty steady at 5mbar for several hours. However, in preparation for the weekend, open 1K Pot needle valve 16th of a turn. |
| 1745 | 1K Pot has settled around 1.73K and P2 reads 65mbar. |
| 1800 | Start 44 hours of 30 min Ba series bias sweeps at +/+1 V to +/+13 V and -/-1 V to -/-13 V, then +/+5V to +/+13V and -/-5V to -/-13V in increments of 1 V with 30s flash and 25min cooldown in between. Triggering on DIB1 Qo at 40ADC bins. 1000 randoms before and after. |
| 1805 | Place Ba source at back of fridge near where top of shield would be for overnight Ba runs. |