1545 After uninstalling the tower, bring it into the clean room for a tower-level electrical check.
Detector G114 DIB 1
side coax ZB1-4-4 date/who: 2016-05-20 DS AK AV
face 5
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 893 Ω (Doing 2→1 gives 1.9kΩ)
2 4 QETA-B 100 to 200 Ω 286 Ω
3 4 QET B 100 to 200 Ω 263 Ω
4 6 QETB-C 100 to 200 Ω 261 Ω
5 6 QET C 100 to 200 Ω 223 Ω
6 8 QETC-D 100 to 200 Ω 222 Ω
7 8 QET D 100 to 200 Ω 166 Ω
2 8 QETA-D 100 to 200 Ω 232 Ω
chassis 9 LED1 Rforward 8 to 9 MΩ open
chassis 9 LED1 Vforward 1.0 to 1.1 V open
9 chassis LED1 Rreverse <1 nS <1 nS
chassis 10 LED2 Rforward 8 to 9 MΩ 710K
chassis 10 LED2 Vforward 1.0 to 1.1 V 1.090 V
10 chassis LED2 Rreverse <1 nS <0.3 nS
1 11 Qo Rbias 24 to 26 nS 26 nS
12 11 Qo C coupling <1 nS <0.5 nS
12 13 Qo Rfeedback 24 to 26 nS 26 nS
1 14 Qi Rbias 24 to 26 nS 26 nS
15 14 Qi C coupling <1 nS 13 nS
15 16 Qi Rfeedback 24 to 26 nS 26 nS
Detector G114 DIB 2
side coax ZB1-4-6 date/who: 2016-05-20 DS AK AV
face 2
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 253 Ω
2 4 QETA-B 100 to 200 Ω 218 Ω
3 4 QET B 100 to 200 Ω 131 Ω
4 6 QETB-C 100 to 200 Ω 224 Ω
5 6 QET C 100 to 200 Ω 204 Ω
6 8 QETC-D 100 to 200 Ω 205 Ω
7 8 QET D 100 to 200 Ω 131 Ω
2 8 QETA-D 100 to 200 Ω 233 Ω
chassis 9 LED1 Rforward 8 to 9 MΩ 725 kΩ
chassis 9 LED1 Vforward 1.0 to 1.1 V 1.091 V
9 chassis LED1 Rreverse <1 nS <0.4 nS
chassis 10 LED2 Rforward 8 to 9 MΩ 718 kΩ
chassis 10 LED2 Vforward 1.0 to 1.1 V 1.091 V
10 chassis LED2 Rreverse <1 nS <0.3 nS
1 11 Qo Rbias 24 to 26 nS 22 nS
12 11 Qo C coupling <1 nS <0.3 nS
12 13 Qo Rfeedback 24 to 26 nS 26 nS
1 14 Qi Rbias 24 to 26 nS 25 nS
15 14 Qi C coupling <1 nS 25 nS
15 16 Qi Rfeedback 24 to 26 nS 26 nS
Detector G114 DIB 3
side coax ZB1-4-8 date/who: 2016-05-20 DS AK AV
face 3
+lead - lead signal typical measured
1 2 QET A 100 to 200 Ω 117 Ω
2 4 QETA-B 100 to 200 Ω 204 Ω
3 4 QET B 100 to 200 Ω 171 Ω
4 6 QETB-C 100 to 200 Ω 155 Ω
5 6 QET C 100 to 200 Ω 114 Ω
6 8 QETC-D 100 to 200 Ω 192 Ω
7 8 QET D 100 to 200 Ω 166 Ω
2 8 QETA-D 100 to 200 Ω 193 Ω
chassis 9 LED1 Rforward 8 to 9 MΩ open
chassis 9 LED1 Vforward 1.0 to 1.1 V open
9 chassis LED1 Rreverse <1 nS open
chassis 10 LED2 Rforward 8 to 9 MΩ open
chassis 10 LED2 Vforward 1.0 to 1.1 V open
10 chassis LED2 Rreverse <1 nS open
Bolded selections are irregular values. May indicate incorrect electrical behavior. No channels are shorted to ground.
1635 Both Qi Coupling Capacitors seem to have little to no resistance in our tower level checkout. Will have to measure the resistance across them directly on their side coaxes to confirm that they are the issue.

Next Day (Run 58)