| | Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50 |
| | Took some Cf series overnight. |
| 0855 | Tower = 31.1 mK, Still = 0.090 mbar, G1/G2 = 150/130 mbar, 1K Pot = 1.55 K / 3.6 mbar, Main bath = 74.8% |
| 0900 | Fill LN2 cold trap Dewar |
| 1505 | Set all QET Biases to 0 on all DIBs. Send 50mVpp 100Hz triangle wave down QET lines for all channels. Triggering on 5000 randoms at 20Hz. This is for prepulse-glitch studies. |
| 1517 | Start 5000 randoms of background data at 0/0 V with triangle waves down all QET lines with FG=5/2/5, OG=3. Stop after randoms are done. |
| 1523 | Power down DCRCs. Remove DSUB pi-filters, attach filter boards to all DIBs with all switches OFF. Repower DCRCs. Zap SQUIDs. |
| 1610 | Perform mini IbIs with shielded pi-filter boards attached and all filter lines switched OFF on all 3 DIBs. |
| 1640 | Remove DCRC33 (DIB2)'s top shield and retake mini IbIs. |
| 1650 | Perform mini IbIs with DCRC33 (DIB2)'s top shield removed, and shielded pi-filter boards attached and all filter lines switched OFF on all 3 DIBs. |
| 1655 | Fill LN2 cold trap full. |
| 1700 | Power off DCRCs. Replace DCRC33's (DIB2)'s top shield. Detach pi-filter boards, reattach DSUB pi-filters, repower DCRCs, zap SQUIDs. |