UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20160308

2016/03/08 (Tuesday)

Previous Day

Setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50
Idle overnight, DCRCs powered
0835 Tower = 31.4 mK, Still = 0.090 mbar, G1/G2 = 125/130 mbar, 1K Pot = 1.54 K / 3.8 mbar, Main bath = 52.0%
0840 Fill LN2 cold trap Dewar
0840 Zap SQUIDs
0900 Use hoist to lower mu-metal shield down below cryostat into the pit. Untied and removed the ropes that normally hold the shield to the cryostat. Wrapped hoist cables in bubble wrap where they contact cryostat or fridge frame to keep fridge electrically isolated.

0915 Set QET biases to standard values and restore all phonon channels so that pulses are seen. 330Hz noise seen in DIB1pC, inverted pulses seen in DIB2pA.
0925 Change DIB2pA QET bias from -112.6 to -120 uA. This gets rid of the upside-down pulses.
1100 While performing mini IbIs on DIBs 1-3 with DSUB pi-filters and mu-metal shield lowered, notice that DIB3 pA was set to the wrong FrontEnd Gain and Driver Gain. While modifying those values, the channel again began experiencing the fat white noise. Power off DCRC3 and let it sit for a minute while the fridge recovers from the related temperature spike.

The TES powers derived from the mini-IbIs are slightly smaller in the no-shield case, by 1.5 ± 1.5 pW. Removing a few outliers (DIB2pA, DIB3pA, DIB3pD) reduces the apparent parasitic power to 0.8 ± 0.6 pW.

1120 Repower DCRC3. Now seeing some high-frequency noise on DIB3, and it's hard to get the Driver Offset set properly to view the traces. Also while using the web-based SQUID settings tool, the values for SQUID Bias and Lockpoint on DIB3 pA were set to duplicated values of DIB3 pB. The issue may not lie with the SQUID Tuning GUI.
1130 Unpower DCRC3 again and let it cool down for longer this time.
1145 Repower DCRC3.
1205 All the channels on DIB3 were railed, but adjusting the Driver Offsets brought them back and there's no fat white noise.
1215 While investigating other channels in preparation for Tc measurements, DIB1 pC has ~50mVpp 330Hz sinusoidal noise on it. This is now present on all DIB1 channels. Unpower all DCRCs and allow fridge to cool (was settled at ~33.5mK which is a little high).
1225 After repowering the boards, and zapping the SQUIDs, the sinusoidal noise is still present on DIB1 channels, and some sharkfin noise is present on DIB2 pB and DIB3 pB. Unpower boards for longer this time.
1310 Repower boards, zap SQUIDs, but the sinusoid won't go away. Maybe it's because the mu-metal shield is down? Time to just bull forward regardless. QET biases have all been set to 0, so send 10mVpp triangle waves down the channels to measure Tc's.
1315 While forging ahead and sending triangle waves down DIB1's QET lines, the fridge temperature spikes up to ~90mK. When investigating DIB3's phonon traces, the fat white noise has shown up again. Unpower DIB3 and allow temperature to settle.
1525 DIB3 high-frequency noise returned, spontaneously it seems, with heating of fridge - temp rose to 60 mK then fell then rose again to 55 mK then settled at 38 mK, clearly elevated.
1530 Tightening the connection between the DSUB connectors for DIB3 railed the phonon channels, but after adjusting the Driver Offsets, we no longer saw the fat white noise. Tighten the connections for all 3 DIBs for good measure. It was also noted that all the SQUID Bias and Lockpoint settings on DIB3 had been set to the same values, and only the web-interface for SQUID setting was used.
1545 All the channels are alive now, though some of them still have uncharacteristically high in magnitude Driver Offsets. And DCRC3 began showing some of the 330Hz sinusoidal noise.
1720 Finish measuring Tc's of all DIBs.
G106a Tc Tower temperature (mK)
heating cooling
Channel s/c→trans trans→normal normal→trans trans→s/c
DIB1 pA 85.5 87 86.5 84.8
DIB1 pB 85.9 87 86.6 84.9
DIB1 pC 85.9 86.3 86.3 85.0
DIB1 pD 83.6 84.4 84.4 83.0
DIB2 pA 82.1 84.6 84.3 81.0
DIB2 pB ? ? 90.0 86.8
DIB2 pC 85.2 87.8 87.8 85.0
DIB2 pD 82.5 83.2 82.8 81.5
DIB3 pA 80.3 83.9 84.0 81.5
DIB3 pB 80.7 82.1 82.2 78.4
DIB3 pC 83.5 85.0 84.9 82.1
DIB3 pD 88.3 88.6 88.4 87.7
No mu-metal shield present
Italicized values are approximate

If we take the Tc to be the average of the s/c→trans and normal→trans values (1st and 3rd columns) and compare the no-shield case to the shield case, we find no significant difference. No-shield Tc values are higher by 0.1 ± 0.8 mK. If we remove four outliers (DIB1pB, DIB2pA, DIB2pC, DIB3pA), the no-shield Tc values are higher by 0.6 ± 0.4 mK. We don't expect the Tc values to be higher with the shield present, if anything they'd be lower.

Comparing the mini-IbIs results to this, there is some evidence that two outlier channels may be susceptible to the lack of a mu-metal shield. DIB2pA and DIB3pA have lower Tc's (-.3/-1.4 mK) and significantly higher parasitic powers (5/4 pW). These are two adjacent quad channels: Side2pB and Side2pC.

1730 Fill LN2 cold trap full.
1735 Commence LHe transfer (34.7%).
1805 Cease LHe transfer (92.1%).

Next Day