2015/11/17 (Tuesday)

Previous Day

Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50.
Overnight: taking Y data, last flash at 7:40.
0845 Tower/MC(H35)/MC sensor/Cold Plate = 31.0/25.6/41.0/154 mK, Still = 1.29 K / 0.12 mbar, G1/G2 = 140/140 mbar, 1K Pot = 1.73 K / 22 mbar, Main bath = 19.0%
0850 Remove Y-88 source from fridge, return to safe
0910 Stop circulation, begin recovering mixture: close V13a, open V9. Now pumping from still to dump.
1010 Remove top corner pieces of shield, store under wood table. Each corner = 4 lead sheets, 8 poly layers, 1 aluminum base with 2 screws
1020 Turn off DCRCs. Surge in Still temperature&pressure. Before turning off DCRCs, Still = 1.03 K / 0.036 mbar; after, Still = 0.83 K / 0.030 mbar
1100 Close large speedivalve to stop pumping on 1K Pot. P1 falls to 0.012 mbar. Close V4a to isolate 1K Pot pump inlet. Divert 1K Pot pump exhaust from recovery to air.
1110 Open V5 to pump directly from condenser and Trap 1 to dump. Turn on MC heater to 10 mW. MC → 1.4 K
1120 Turn on Still heater to 10 mW. Still → 5.3 K
1430 Dump pressure (G2) flat at 810 mbar, MC rises to 8.1 K.
1435 Finish removing shield from fridge.
1445 Close still valve and condenser valve for a few minutes to thoroughly pump out Trap 1.
1450 Close 12a and associated speedivalve to isolate Trap 1. Reopen still and condenser valves.
1455 Close still and condenser valves. Dump (G2) = 810.2 mbar. P_atm = 29.55 “Hg so Dump = 797.7 mbar. Mixture recovered.
1500 Add 4.6 mbar He exchange gas to IVC. 1K Pot temperature rises then falls; MC temperature falls to 4 K (since it is now in thermal contact with IVC).
1505 Close valves to isolate IVC. Then changed my mind - reopened IVC to gauge to monitor pressure.
1510 Close V6, V1, V5. Only V9 open. Turn off 3He pump.
1515 Turn off MC heater and Still heater (probably won't help much to boil off the bath)
1515 Main bath = 10.6%
1640 Close valves to recovery system. Lower cryostat Dewar (main bath) about 6”. Use clean tent HEPA filter fan to blow on exposed part of fridge to reduce condensation.
1705 MC = 4.4 K (Cernox), Sorb = 17.5 K, IVC = 7.4 mbar