2015/11/01 (Sunday)

Previous Day

Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50.
Overnight: Taking Cf data. Last flash at 22:50 last night. Tower temp stable.
0755 Main bath = 30.6%. Note: Daylight Saving Time has ended.
0801 Start LHe transfer at 29.7%
0815 Tower/MC(H35)/MC sensor/Cold Plate = 31.0/26.9/412.2/162 mK, Still = 1.43 K / 0.11 mbar, G1/G2 = 125/125 mbar, 1K Pot = 1.69 K / 14 mbar
0815 Remove Soudan Cf source from fridge, put back in safe.
0820 Fill LN2 cold trap Dewar
0844 Finish LHe transfer at 90.7%
0850 Start 20 cycles of 30min runs of background data at +/-2V, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after. Using Wednesday's thresholds.
1752 For some reason the DAQ hung up on the _1752 series. Only 9 series taken.