| 1010 | Tower/MC(H35)/MC sensor/Cold Plate = 31.0/25.3/43/158 mK, Still = 1.38 K / 0.085 mbar, G1/G2 = 155 mbar, 1K Pot = 1.72 K / 22 mbar, Main bath = 63.1%. |
| 1015 | Power cycle DCRC55 because of pB 80Hz noise. |
| 1028 | Channels on DIB1 set up, and no sign of 80Hz noise. Wait and see how long it takes to come back. |
| 1100 | Fill LN2 cold trap full. |
| 1110 | Still no 80Hz noise. Mini bake detector and see if it comes back after that. |
| 1155 | Still no 80Hz noise. Take background data, test for bad channel: Start DAQDAQ2 in same configuration as yesterday, except only 1 cycle of -/+2V and -/+4V, and OR trigger now includes DCRC55 pB again. |
| | Through randoms, and no sign of 80Hz noise. |
| | Notice that timing link is not set up, so we are able to compare it with the potentially unsynchronized data from Tuesday to verify whether that may have been the cause. |
| 1158 | Establish timing link. |
| 1440 | After data series finished, still no 80Hz noise. |
| 1505 | Take background data, test for bad channel: Start another run of DAQDAQ2 with same DCRC settings as before. Run details: 5min series of -/+2V then -/+4V data with 6sec of flashing and 15min cooldown in between. |
| 1530 | Unexpected error after 2nd series. It appears some issue with triggering, so -/+4V series may have been cut short. After hitting “continue” MultiDAQ proceeded to flash. |
| 1545 | Still no 80Hz noise even after more data taking. |
| 1733 | Fill LN2 cold trap full. |
| 1810 | After fridge cooled after some MIDAS tests, attempt to take overnight background data. The channels are all railed and some won't un-rail even with changing the offsets. Confused. Unpower boards overnight. |