| | Current setup: DIB1/2/3 → cables → 3.2 MHz filters → HV1/2 → DCRC50/55. Unused LED is grounded. |
| | Overnight, Cf source in shield activating detector, detector baking (2 LEDs), Tower = 284 mK, Still = 0.10 mbar |
| 0930 | Remove Cf-252 source from fridge. |
| 0930 | Fill LN2 cold trap full. |
| 0930 | Stop baking detector. Tower cools to 32 mK after 20 minutes |
| 1132 | Commence LHe transfer (17.9%). |
| 1210 | End LHe transfer (78.9%). |
| 1315 | Tower/MC(H35)/MC sensor/Cold Plate = 30.8/24.3/39.7/150 mK, Still = 1.30 K / 0.080 mbar, G1/G2 = 155 mbar, 1K Pot = 1.72 K / 24 mbar, Main bath = 78.5%. |
| 1400 | Charge channels have the regular spike pattern seen before so we will need to power cycle the DCRCs. Also some sort of triangle waveform in HV1pB. |
| 1445 | DCRCs repowered and SQUIDs zapped, but now a 250 Hz sine wave dominating the phonons. Power cycle again. |
| 1540 | Channels look good now except HV1pB, which has some very regular pattern at 80 Hz, not the same pattern as before but probably the same frequency. Looks for all the world like response to some sort of test signal, no idea. |
| 1600 | Preparing to take data with MultiDAQ at a few HV settings. The purpose is to look for “effective Luke voltage” in phonon spectra and see how noise changes with HV, inspired by UCB studies. |
| 1620 | After some struggle with MultiDAQ choose the following settings: |