0945 Filled LN2 cold trap full. LHe 36.7%, G1/G2 = 160/144 mbar, 1K pot = 1.742K, Tower = 31.5mK.
0947 Reversed QET bias polarity on DIB2, right-side-up pulses! Signals look good on all 3 DIBs.
1008 DCRC55 PB developed a triangle wave despite what the appropriate registers said and did not respond to “stop test signal” requests. Power cycling and zapping squids fixed this.
1042 DIB1 DCRC55 squid settings were incorrect. David loaded some saved from the previous run (Run 53) which are listed below. DIB2/3 settings remain the same as yesterday.
DIB1 DCRC55
Front End Gain = 1, Output Gain = 0
Squid Bias (uA) Lockpoint (mV)
PA -87 -4.86
PB -86 -3.89
PC -84 -4.28
PD -84 -2.82
1045 Run IbIs
1305 Started LHe transfer (31.9%)
1335 Cease LHe transfer (91.3%).
1620 Fill LN2 cold trap full.