2015/06/29 (Monday)

previous day

0830 Remove Cf source from fridge (was on over the weekend to activate detector)
0836 Fill LN2 cold trap Dewar
0842 Stop flashing detector. Note that tower temperature only reached 128 mK during flashing. Turned off the three DIBs in sequence: DIB1 off, 128 mK → 44 mK. DIB2 off, no temperature change. DIB3 off, 44 mK → 34 mK.
0845 Experimented with DIB2 flashing. Noticed in the PuTTY session that during flashing rd 30 returned 0003, thus FETs were off. Tried flashing from DetectorControl - no temperature change. Tried wr 30 0003 - no temperature change. Tried wr 30 0803 - temperature shoots up >90 mK. Conclusion: at least on DCRC33, LEDs do not turn on unless FETs also turned on.
0900 Reduce thresholds to lowest values possible without triggering frequently on noise:
Current phonon threshold settings
pA pB pC pD
DIB1 15 20 15 15
DIB2 20 15 20 15
DIB3 15 15 10 10
Current FrontendGain/OutputGain(setting)/DriverOffset settings
pA pB pC pD
DIB1 7/1/-0.018 7/1/-0.014 7/1/-0.036 7/1/-0.024
DIB2 7/1/0.034 4/1/0.044 7/1/0.032 7/1/0.030
DIB3 7/1/-0.086 7/1/-0.022 7/0/-0.108 7/0/-0.156
0913 Set to -4 / +4 V and start bkg data run
0921 Tower = 31 mK, Still = 2 mW / 0.11 mbar, G1/G2 = 215/210 mbar, 1K Pot = 1.68 K / 15 mbar, Main bath = 36.8%
0930 Set to 0 / 0 V and flash detectors (all 3 DIBs) for 10 seconds. Temperature peaks around 150 mK
0945 Tower did not cool all the way, only to 42 mK. Had to turn off DIB3 FETs to make it cool to base.
0952 Tower = 31 mK. Set -4/+4 V. Start bkg data run
1008 Stop data, 0/0 V, flash 15 sec. (Note: for DCRC50→DIB3 we are now doing “wr 30 0803” to flash and “wr 30 0” to stop flashing, i.e. turning FETs on and off along with LEDs.)
1018 With 15 second flash, Tower → 200 mK and it takes 10 minutes to cool below 32 mK
1022 Set to -4/+4 V. Start bkg data run
1039 Stop data, 0/0 V, flash 30 sec.
1051 With 30 second flash, Tower → 242 mK and it takes 12 minutes to cool below 32 mK
1055 Set to -4/+4 V. Start bkg data run
1118 Stop data, 0/0 V, flash 30 sec.
1118 With 30 second flash, Tower → 225 mK and it takes 12 minutes to cool below 32 mK
1137 Set to -4/+4 V. Start bkg data run
1154 Stop data, 0.0 V, flash 30 sec. Tower peaks at 243 mK.
1215 Power down all boards, remove DCRC55 for voltage calibration.
1400 Power on boards. Zap SQUIDs. Set all SQUID/QET settings to previous values.
1545 Test flashing with DetectorControl vi. LEDs from all three DIBs work (produce temperature increases).
1549 Test flashing with PuTTY. Now DIB3 works the same as it used to: you don't have to turn off the FETs to get the LEDs to turn off.
1616 Several phonons flatline but recovered with relocks. All settings still good with one exception:
Updated DriverOffset settings
pB
DIB1 -0.094
1627 Flash for 30 seconds
1641 Set to -4/+4 V. Start bkg data run. (Note that we are now using properly calibrated bias for DCRC55→DIB1, but it's very close to the nominal setting: to get -4 V in Q1/Q2 requires nominal settings of -3.999/-3.965 V)
1520 Checked pressure of LHe transfer tube vacuum jacket with external pressure sensor; jacket pressure at 2.5E-2 mbar (including the volume of 3 bellows). Q compound appears to be successfully reducing the leak, though it appears to have been smudged during the previous transfer. Cover Q compound with duct tape to prevent future movement.
1535 Commence LHe transfer (25.5%).
1415 End LHe transfer (91.7%).
1730 Fill LN2 cold trap full.
Attach 3.2 MHz Pi Filter to DCRC 55 (DIB1) and 10 MHz C Filter to DCRC 33 (DIB2). Continue studies tomorrow.

next day