2013/08/12
| 1925 | Commence tower assembly procedure. |
|---|---|
| 2010 | Mount Am-241 source facing Side 1 Phonon Channel B (A1). |
| Remove G106a from purge cabinet to mount on tower. | |
| 2225 | Perform tower level continuity checks: |
| Detector | G106a | DIB | 1 | |
|---|---|---|---|---|
| side coax | ZB1-4-4 | date/who: | 2013-08-12 | HC |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 126 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 206 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 192 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 211 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 188 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 201 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 155 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 180 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 775 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.092 V |
| 9 | chassis | LED1 Rreverse | <1 nS | ≈ 1 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 752 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.093 V |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.3 nS |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 26 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.2 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.2 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G106a | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-4-6 | date/who: | 2013-08-12 | HC |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 201 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 186 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 124 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 183 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 168 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 174 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 123 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 196 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 774 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.092 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.2 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 741 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.089 V |
| 10 | chassis | LED2 Rreverse | <1 nS | <9 nS |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 25 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.2 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 25 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.2 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G106a | DIB | 3 | |
|---|---|---|---|---|
| side coax | ZB1-4-8 | date/who: | 2012-08-12 | HC |
| face | 6 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 180 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 224 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 204 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 167 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 153 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 167 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 148 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 195 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 769 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.087 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <11 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 753 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.086 V |
| 10 | chassis | LED2 Rreverse | <1 nS | OL |
| 2255 | With the exception of three LEDs (LED2 on DIB2 and LED1 & LED2 on DIB3) no shorts to ground found. State of aforementioned LEDs appears to be worse than that recorded during last checkout at start of Run 33. Three LEDs should be enough to neutralise the crystal. |
|---|---|
| 2330 | Tower assembly procedure complete. Bag up tower for installation tomorrow. |