2012/11/27
| G106a arrived today. Pressure in puck was 7 psig. Puck was transferred to clean room and the detector was removed. A detector level check was performed. |
| Note that G106a has the same charge and phonon channel configuration as G103a. However, the number of LEDs that G106a has is 6. |
| Table 1: Detector level continuity check. | |||
|---|---|---|---|
| DIB 1 | DIB 2 | DIB 3 | |
| Signal | Measured | Measured | Measured |
| QET A | 31.2 Ω | 117.3 Ω | 99.3 Ω |
| QET B | 105.1 Ω | 31.2 Ω | 128.3 Ω |
| QET C | 94.5 Ω | 77.1 Ω | 71.4 Ω |
| QET D | 59.1 Ω | 30.2 Ω Ω | 62.8 Ω |
| LED1 R forward | 793 kΩ | 754 kΩ | 783 kΩ |
| LED1 V forward | 1.05 V | 1.044 V | 1.05 V |
| LED1 G backward | <0.03 nS | <0.1 nS | <2.6 nS |
| LED2 R forward | 768 kΩ | 754 kΩ | 776 kΩ |
| LED2 V forward | 1.051 V | 1.049 V | 1.049 V |
| LED2 G backward | <0.3 nS | <9 nS | Open |
| Qo(+ve)-Qi(-ve) | 67.3 Ω | 58.1 Ω | N/A |
| Qi(+ve)-Qo(-ve) | 64.2 Ω | 52 Ω | N/A |
| LED 2 conductance for Dib 2, and LED 1 for Dib 3 are a bit high. LED 2 on Dib 3 is fully conducting. Not a good sign. |
| 2120 | Finished tower assembly and checkout. |
|---|---|
| Source is located next to channel C (C1). |
| Table 2 - Tower checkout | ||||
|---|---|---|---|---|
| Detector | G106a | DIB | 1 | |
| side coax | ZB1-4-6 | date/who: | 2012-11-27 | RR & AK |
| face | 2 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 126 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 205 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 192 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 210 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 187 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 198 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 155 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 177 Ω |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 756 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.094 V |
| 10 | chassis | LED2 Rreverse | <1 nS | <0.7 nS |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 784 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | not measured |
| 9 | chassis | LED1 Rreverse | <1 nS | not measured |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 26 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.3 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 25 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.5 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G106a | DIB | 2 | |
|---|---|---|---|---|
| side coax | ZB1-4-4 | date/who: | 2012-11-27 | RR & AK |
| face | 5 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 203 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 186 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 124 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 183 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 169 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 175 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 125 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 198 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 782 kΩ |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 744 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.094 V |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.09 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.4 nS |
| 10 | chassis | LED2 Rreverse | <1 nS | <5 nS |
| 1 | 11 | Qo Rbias | 24 to 26 nS | 25 nS |
| 12 | 11 | Qo C coupling | <1 nS | <0.2 nS |
| 12 | 13 | Qo Rfeedback | 24 to 26 nS | 26 nS |
| 1 | 14 | Qi Rbias | 24 to 26 nS | 26 nS |
| 15 | 14 | Qi C coupling | <1 nS | <0.2 nS |
| 15 | 16 | Qi Rfeedback | 24 to 26 nS | 26 nS |
| Detector | G106a | DIB | 3 | |
|---|---|---|---|---|
| side coax | ZB1-4-8 | date/who: | 2012-11-27 | RR & AK |
| face | 6 | |||
| +lead | - lead | signal | typical | measured |
| 1 | 2 | QET A | 100 to 200 Ω | 181 Ω |
| 2 | 4 | QETA-B | 100 to 200 Ω | 224 Ω |
| 3 | 4 | QET B | 100 to 200 Ω | 205 Ω |
| 4 | 6 | QETB-C | 100 to 200 Ω | 167 Ω |
| 5 | 6 | QET C | 100 to 200 Ω | 153 Ω |
| 6 | 8 | QETC-D | 100 to 200 Ω | 166 Ω |
| 7 | 8 | QET D | 100 to 200 Ω | 148 Ω |
| 2 | 8 | QETA-D | 100 to 200 Ω | 193 Ω |
| chassis | 9 | LED1 Rforward | 8 to 9 MΩ | 774 kΩ |
| chassis | 9 | LED1 Vforward | 1.0 to 1.1 V | 1.089 V |
| 9 | chassis | LED1 Rreverse | <1 nS | <0.9 nS |
| chassis | 10 | LED2 Rforward | 8 to 9 MΩ | 765 kΩ |
| chassis | 10 | LED2 Vforward | 1.0 to 1.1 V | 1.1 V |
| 10 | chassis | LED2 Rreverse | <1 nS | 3.2 nS |
| LED 2 on Dib 2 is still high. Now LED 2 on Dib 3 is as well, while LED 1 is okay. Also, QET A on Dib 1 is lower than expected. |