2010/02/22

Time
15:50 Zero DCRC according to the calibration table, zapping all SQUIDS.
16:03 Zapping all SQUIDS again.
16:18 Turn on test signal slope = 43, mag.=17606, tune SQUID.
Turn off test signal, zap A again.
Tune SQUIDs
SQUID settings
Channel SQUID Bias (uA) Lock point
A 90 3.5
B 100 3.4
C 92.16 3.2
D 95 3.5
SQUID modulation curves
Closed loop settings
Channel Frontend gain Polarity
A 3 Non-inverting
B 3 Non-inverting
C 3 Non-inverting
D 3 Non-inverting
Closed loop response
Time
17:44 Save current settings as Preset 2. Need solder a pin on Feedback B, do it later.
18:28 Tower thermometer ~ 2min after MC at base T. MC temperature fluctuates. Set still power to 5mW
18:37 MC temperature fluctuates, Set still power to 7mW
18:48 MC temperature fluctuates, Set still power to 8mW
19:07 MC temperature fluctuates, Set still power to 0mW
19:34 MC temperature drops to 26mK
19:59 MC temperature stays at 25.9mK
Measure Ic as a function of temperature. Didn't get meaningful data because temperature fluctuation.
Measure Tc.
22:00 Fill N2 cold trap full

Ic vs T

Ic in the table below is the function generator output, or the input to the QET bias line.

MC T (mK) Tower T (mK) Ic (mVpp) MC heater (uW)
phA phD
25.7 124.15 kΩ 91 196 0

Tc

Send 1mVpp 100Hz triangle signal, apply different power to the mixing chamber, and watch at which resistance (temperature) the triangle wave disappear or reappear.

MC power (uW) 50 0 20 0 10 0 10 6 5
Temperature change ↑ ↓ ↑ ↓ ↑ ↓ ↑ ↑ ↓
phA (kΩ) missed 26.96 31.6 26.9 29.6 27.67 28.8 29.4 28
phA (mK) missed 81.3 73.8 81.4 76.7 80.0 78.0 77.0 79.4
phD (kΩ) missed 26.75 30.8 26.4 28.8 27.6 28.1 28.5 27.7
phD (mK) missed 81.7 74.9 82.4 78.0 80.1 79.2 78.5 79.9

From the above table, Tc of A is between 75mK and 80mK and Tc of D is between 80mK and 85mK. Numbers from CWRU