UMN CDMS Wiki archive
start › cdms › k100 › group_meetings › 20240312
Teleconference link: https://umn.zoom.us/j/9024497080?omn=91370498825
Meeting ID: 902 449 7080
Domestic: 1-408-638-0968 (US toll)
Access code:
  • Priority Updates
  • General Updates
    • Himangshu
    • Shubham
    • Zach
  • Fun Update