  • Lab Update - Run 77 preparation
    • Detector Control GUI working again after recompiling and defining appropriate variables.
    • Found the short on the Source mover coil. The coil wires were loosely held at Pin1 and Pin 4 which came off when I removed the Kapton tape. Resolder the pins and did a quick test to make sure the Y-arm is moving smoothly.
    • Zach made the washer plate for the Am source stage using a copper piece. Pictures
    • Zach started cryocleaning yesterday. We are planning to start cooldown on Monday.
    • Calibration curve for the new thermometer implemented in the Labview VI.