UMN CDMS Wiki archive
start › cdms › k100 › group_meetings › 20221229
Teleconference link: https://umn.zoom.us/j/2461438631
Domestic: 1-646-558-8656 (US toll)
Domestic: 1-408-638-0968 (US toll)
Access code: 908-227-5310
  • Lab Update - Run 76 post-mortem
    • Closed IVC to leak-check stripline elbow, and other seals - no leaks (at room temperature)
    • repairing tower wires:
      • turns out all wires broke between 4K and 600mK solder pads - all still intact between 600mK and 30mK
      • Existing wires are actually similar to motor wires (copper-clad NbTi), but the copper is etched away between solder pads leaving only superconductor
      • Copper is a bad idea in terms of heat leak
      • Replaced the wires with manganin, a low-thermal-conductivity wire often used in cryo
      • but this has disadvantages - the resistance of one wire between 4K and 600mK is almost 1 Ohm. Unsuitable for phonon readout - but this tower face has other problems so we don't use it for phonon channels.
    • Electrical checkout of detector, looking for clues about HV leakage problem
      • Resistances between channels are smaller. Biggest difference is between Side1 channels (note Side1 was the HV side when the HV mishap occurred)
      • Possibly potential differences occurred between channels, causing arcing between nearby rails and creating conducting paths
      • It's conceivable that when leakage is significant, ordinary biasing doesn't quite bias all channels equally. Then eventually maybe more leakage starts occurring between Side1 channels
      • This wouldn't really explain why biasing Side2 also seemed to show extra leakage above ~ 100V. But the leakage looked a little different.
      • Likely the leakage problem is at the detector itself (rather than SQUETs or DIBs or something). Possibly detector “surgery” could help, but it would probably be hard to know where to do it.
    • Inspect damaged SQUETS. Visual inspection doesn't show any broken bonds. Damage could be in chip.
      • Potentially we could swap out chips to get four good ones on a SQUET. But you can't wirebond them too often, the bond pads get covered
    • Began modifying other SQUETs to reduce shunt resistances
      • Bonder in Hanany lab (note: Vlad Pribiag is currently in charge of the bonder) needs repairs
      • Paul Crowell hooked us up with the Ke Wang lab, they have their own bonder. Wei Ren let us use it under supervision.
  • Priority Updates
  • General Updates
    • Himangshu
    • Shubham
    • Jack
    • Derek
    • Zach
    • Scott
    • Giulia
  • Fun Update