  • Lab Update
    • Testing Side2-readout HV board
      • noise issues mostly gone, still a little more high-frequency noise in DCRC1 than before
      • best noise width seen was 5.2 eVee, with source over pC. Somewhat worse noise width with source at center.
      • Trying lower points in transition; with dP/dI reduced from 1.0 uV to 0.9 uV see some improvement in noise width
      • Strong pulse-widening seen, makes calibrating more complicated: example plots
        • Side2 Tc's are lower; this seems to make “saturation” effects stronger
    • Plan to end Run75 this week
  • Priority Updates
  • General Updates
    • Himangshu
      • Fridge Diagram Complete. Plan to install code on iPC, sometime this week.
      • Fridge Alarm system notification on slack channel started. Working with backend python code, starting from NEXUS code NEXUS Alarm code
      • Installed CDMSBats on Mac. This is an initial effort to use updated CDMSBats with our lab data.