- Lab Update
- Source mover performance
- We can send any number of steps to the motor without the coil resistance going up
- Heating is still significant
- Can move 5 mm/hour and fridge remains stable (c/w 1 mm/hour in Run71). May be able to go a little higher
- We've tried as high as 13 mm/hour but above 9 mm/hour the fridge can't keep up - not sure what would happen if we did that for more than a few mm
- ~3 kHz noise
- A couple of charge channels showed high noise with pickup in the others, too high to see 60keV line
- As of yesterday this noise is 10x smaller - so we can take meaningful data again (if it doesn't come back)