| XXXX | XX/XX/XX, XX:XX | 1 hour by 168 | Ti foil roll from SLAC | 11337 | Screening of sample. |
| 11195 | 02/13/13, 12:35 | 1 hour by 168 | VESPEL SCP-5050 from UCB | 10497 | Screening of sample. |
| 10422 | 01/15/13, 09:47 | 1 hour by 168 | Derlin Bobine and Rollers for SuperNEMO | 9826 | Screening of sample. |
| 9819 | 12/17/12, 12:46 | 1 hour by 168 | PMMA Bobine for SuperNEMO | 9321 | Screening of sample. |
| 8986 | 11/26/12, 13:41 | 1 hour by 168 | N/A | 8985 | Testing of DAQ after hardware error and sytem reboot. |
| 8799 | 11/21/12, 15:00 | 1 hour by 168 | Noryl blocks | 8145 | Screening of sample. |
| 7998 | 10/24/12, 14:27 | 1 hour by 168 | 2020 Graphite, Unpurified | 7809 | Screening of sample. |
| 7739 | 10/16/12, 13:30 | 1 hour by 168 | Screening, 4.7 nF Capacitors | 7473 | Screening of sample. |
| 7376 | 09/28/12, 09:46 | 1 hour by 168 | Screening, UF-4S Purified | 6801 | Screening of sample. Unsure whether file 6465 is with background or sample so remove from analyses. Ran until Gopher dry again, files 7172-7183 are from warm-up. Ran after cool down. |
| 6297 | 09/13/12, 10:00 | 1 hour by 168 | Testing, Background(0912) with 1 Poly Brick | 6129 | Testing of Gopher after the LN2 ran dry to see if any major changes. |
| 5983 | 09/04/12, 09:27 | Background, 1h | Warm-Up | 5967 | Gopher's LN2 ran dry during the the background run. The files here are from when Gopher was warm and could not record any data. They are BAD and should not be used. |
| 5966 | 09/03/12, 15:41 | Background, 1h | Background(0812) with 1 Poly Brick | 5289 | Running 168 1 hr files, repeated. One week long total and 1 hr allows for minimal loss from autofill. Background run after hot ZrO2 sample. Data taking not stopped fully between changing. Files 5168,5169 are from the change and bad. Accidentally, data was not taken for the weekend of Aug. 11-12. |
| 5167 | 08/03/12, 15:30 | screening, 1h | Zirconium Oxide | 4953 | Running 168 1 hr files. One week long total and 1 hr allows for minimal loss from autofill. 4953 is short and thrown out. |
| 4952 | 08/01/12, 14:18 | screening, 1h | Zirconium Oxide | 4785 | Running 168 1 hr files. One week long total and 1 hr allows for minimal loss from autofill |
| 4784 | 07/25/12, 12:12 | screening, 1h | Zirconium Oxide | 4617 | Running 168 1 hr files. One week long total and 1 hr allows for minimal loss from autofill |
| 4266 | 07/18/12, 11:28 | screening, 12h | Zirconium Oxide | 4249 | Same mode as previous, 12 hr with extra for filling. 4249 File has bad start time |
| 4237 | 07/17/12, 11:24 | screening, 12h | 2020 Graphite Purified | 4185 | 12h screening mode with 2 extra for filling, some short files from filling |
| 4169 | 06/21/12, 15:19 | screening, 2h | Steel After Purification (Betacage) | 3817 | Changed back to 2h run, G2K cleared properly |
| 3810 | 06/03/12, 23:53 | screening, 12h | Steel After Purification (Betacage) | 3803 | Changed to 12h run but G2K was not cleared properly so is apart. of the previous 2h sequence. 3803 is 2+12=14h long. |
| 3721 | 05/30/12, 21:52 | screening, 2h | Steel After Purification (Betacage) | 3719 | New Sample. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 3685 | 05/30/12, 09:21 | screening, 2h | Steel Before Purification (Betacage) | 2781 | Three runs combined. Discard 3705, G2K was left counting while sample was changed |
| 2319 | 05/16/12, 14:57 | screening, 12h | Steel Before Purification (Betacage) | 2307 | |
| 2306 | 05/09/12, 16:25 | screening, 5m by hand | Steel Before Purification (Betacage) 2283 | Autofill run | |
| 2270 | 05/09/12, 13:30 | screening, 12h | Steel Before Purification (Betacage) | 2269 | New Sample. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 2262 | 05/08/12, 12:13 | screening, 12h | Carbon Rods from UCB | 2227 | |
| 2215 | 04/18/12, 16:29 | screening, 80m | Carbon Rods from UCB | 2215 | Failed Autofill test, must rename 2215 → 2203 and use with 2201, 2202 in analysis |
| 2202 | 04/18/12, 15:10 | screening, 12h | Carbon Rods from UCB | 2201 | |
| 2162 | 04/17/12, 15:30 | screening, 12h | Carbon Rods from UCB | 2137 | New Sample. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 2131 | 04/04/12, 13:26 | screening, 12h | Purified AXM-5Q Graphite | 2123 | Auto purge work completed, counted over the weekend. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 2109 | 03/27/12, 01:22 | screening, 12h | Purified AXM-5Q Graphite | 2109 | 1 Day run (12h) run, turned off at night for mine crew to continue purge system work in the morning. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 2108 | 03/24/12, 14:47 | screening, 12h | Purified AXM-5Q Graphite | 2095 | HV Turned back on, run continued. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 2094 | 03/18/12 | screening, 12h | Purified AXM-5Q Graphite | 2087 | Files continued to be saved during purge bag installation while the HV was off. Also 2086 is not 12h long, and should probably be excluded along with 2087-2094 from analysis. The CNF files were deleted but as the following files (2095-2108) are part of the same series, they (2087-2094) must be included in the radon script and then removed in the final analysis script. The .txt files can be found in users\matt\data\all_converted |
| 2086 | 03/13/12, 10:03 | screening, 12h for 21 days | Purified AXM-5Q Graphite | 2067 | When restarting the run the spectrum in G2K was accidentally cleared. Run was “stopped” for purge work. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 2066 | 03/03/12, 02:35 | screening, 12h | Purified AXM-5Q Graphite | 2053 | The computer's clock and date were off for this run. G2K put all the data for the first 3 12 hour increments into 2053 and extended the run by 2.5 days. The radon was bad for these files as well so they should not be used at all in the analysis of the graphite cube axm pure sample. Used the Ba133_Co60_16K_100719.CAL calibration file |
| 2045 | 02/09/12, 03:37 | screening, 12h Repeated | Background Run | 1997 | Used the Ba133_Co60_16K_100719.CAL calibration file, Gopher down for Computer updates related to gas system in between 1985 and 1986, UPS and Computer Crash ended run |
| 1985 | 01/11/12, 14:34 | screening, 12h Repeated | Background Run | 1885 | Used the Ba133_Co60_16K_100719.CAL calibration file |
| 1870 | 11/10/11, 20:53 | screening, 12h Repeated | CDMS Pb Arc | 1843 | Used the Ba133_Co60_16K_100719.CAL calibration file |
| 1829 | 10/27/11, 13:43 | screening, 12h Repeated | CDMS Shield poly arcs | 1745 | Used the Ba133_Co60_16K_100719.CAL calibration file |
| 1732 | 09/06/11, 12:21 | screening, 12h Repeated | Background Run, Empty Sample Chamber, New Shield | 1703 | Used the Ba133_Co60_16K_100719.CAL calibration file |
| 1701 | 06/29/11, 21:59 | screening, 12h | Background Run, Empty Sample Chamber | 1689 | |
| 1685 | 06/23/11, 02:33 | screening, 12h | Background Run, 2 Poly Bricks | 1647 | |
| 1627 | 03/17/11, 17:19 | screening, 12h | Graphite Cube “ACF-10Q” from Miguel (UCB) | 1619 | Run cut short by fire |
| 1608 | 03/06/11, 16:51 | same | same | 1605 | |
| 1604 | 03/03/11, 14:38 | same | same | 1591 | |
| 1590 | 02/24/11, 14:00 | screening, 12h | Graphite Cube “AXM-5Q” from Miguel (UCB) | 1577 | |
| 1578 | 02/16/11, 17:05 | same | same | 1563 | |
| 1562 | 02/09/11, 16:59 | screening, 12h | Graphite Tube from Miguel (UCB) | 1549 | |
| 1508 | 01/14/11, 12:46 | same | same | 1507 | |
| 1506 | 01/13/11, 12:35 | same | same | 1493 | |
| 1492 | 12/27/10, 15:53 | same | same | 1479 | |
| 1476 | 12/20/10, 15:46 | screening, 12h | Ti sheet from Miguel (UCB) | 1465 | |
| 1457 | 12/14/10, 02:27 | Screening, 12h | Graphite Cube ZXF-5Q | 1451 | |
| 1450 | 12/10/10, 14:13 | ? | Probably Graphite Cube ZXF-5Q | 1450 | Not used in the ZXF-5Q analysis |
| 1449 | 12/10/10, 06:03 | Screening, 12h | Graphite Cube ZXF-5Q | 1437 | |
| 1434 | 12/03/10, 11:52 | screening, 12h | background counting in 12-hours chunks immediately after the door was closed. LN purge flow rate is at 3.5 SCFH. | 1429 | |
| 1428 | 11/30/10, 11:05 | same | same | 1415 | |
| 1414 | 11/23/10, 09:38 | screening, 12h | same | 1401 | |
| 1400 | 11/16/10, 06:11 | screening, 4h | inserted SP-22 from Miguel | 1397 | |
| 1360 | 10/29/10, 14:31 | screening, 4h | Miguel's brown sample. 2“ diameter by 9.5” long, 719 grams | 1357 | |
| 1343 | 10/26/10, 01:00 | same | same | 1329 | good data =1343+1328+1309+1300 -1297 |
| 1328 | 10/14/10, 12:09 | same | same | 1315 | |
| 1309 | 10/05/10, 16:26 | screening, 12h | same | 1301 | |
| 1300 | 10/01/10, 09:04 | screening, 4h | Empty cavity. Background counting | 1297 | |
| 1284 | 09/00/10, 00:00 | 4 Lead Bricks | Screening, 12h | 1251 | Contains many sub-runs clumped together |
| 1243 | 09/14/10, 05:02 | same | same | 1237 | Probably still the caps, but there is a time large gap between 1236 and 1237 |
| 1236 | 08/31/10, 16:23 | same | same | 1223 | |
| 1222 | 08/24/10, 13:58 | same | same | 1209 | 1222 has a lot of Rn-222 |
| 1208 | 08/17/10, 11:35 | Caps | Screening, 12h | 1195 | Definitely |
| 1194 | 07/29/10, 10:11 | Caps | Screening, 4h | 1191 | Probably |
| 1178 | 07/00/10, 00:00 | Unknown | Unkown | 1163 | |
| 1162 | 07/19/10, 10:20 | screening, 12h | W scrap (rods, 2 plates) with unknown origin from UMN shop. Total weight of screened W was 1455 grams | 1149 | |
| 1134 | 07/09/10, 15:45 | same | same | 1131 | good data =1134+1130+1116+1102 +1088+1074+1059-1047 |
| 1130 | 07/02/10, 17:36 | same | same | 1117 | |
| 1116 | 06/25/10, 16:58 | same | same | 1103 | |
| 1102 | 06/16/10, 12:44 | same | same | 1089 | |
| 1088 | 05/28/10, 16:54 | same | same | 1075 | |
| 1074 | 05/21/10, 16:17 | same | same | 1061 | |
| 1059 | 05/14/10, 16:06 | screening, 12h | same | 1047 | |
| 1046 | 05/08/10, 08:36 | screening, 4h | 8 SQUETs atop the poly brick pedestal | 1043 | |
| 1036 | 05/07/10, 12:26 | same | same | 1029 | good data =1036+1019+1014 +1000+986+972-969 |
| 1019 | 04/26/10, 03:07 | same | same | 1015 | |
| 1014 | 04/23/10, 13:53 | same | same | 1001 | |
| 1000 | 04/16/10, 13:18 | same | same | 987 | |
| 986 | 04/09/10, 11:49 | screening, 12h | same | 973 | |
| 972 | 04/02/10, 04:39 | same | same | 969 | |
| 968 | 04/01/10, 09:37 | screening, 4 h | with inserted 8 (3500 grams) poly bricks from CDMS inner layer, without two supporting polybricks | 965 | |
| 962 | 03/31/10, 15:50 | screening, 12 h | same | 951 | good data =962+950+906+882+852-849 |
| 950 | 03/24/10, 18:07 | screening, 12 h | same | 937 | |
| 906 | 03/15/10, 14:22 | screening, 12 h | same | 895 | |
| 882 | 03/06/10, 11:50 | screening, 12 h | same | 853 | |
| 852 | 02/19/10, 11:16 | screening, 4 h | with inserted one of the lead bricks from CDMS's outer shield layer | 849 | |
| 841 | 02/17/10, 14:06 | same | same | 837 | good data =841+836+828-825 |
| 837 | 02/16/10, 01:29 | same | same | first | |
| 836 | 02/15/10, 13:18 | same | same | 829 | |
| 829 | 02/12/10, 01:18 | screening, 12 h | same | first | |
| 828 | 02/11/10, 10:08 | same | same | 825 | |
| 825 | 02/10/10, 22:08 | screening, 4 h | CMS PbW crystals: one 1116 grams, the other 1112 | first | |
| 804 | 02/10/10, 12:37 | screening, 12 h | single stick of Noryl (1.25×1.5×8“ weighing 300g) for beta cage on top of 2 poly bricks | 783 | good data =804+782-779 |
| 778 | | background | with 2 milled poly bricks to be used as mid-size-sample support | | good data =778+770+762+720-717 |
| 699 | | background | after inner Cu box was washed in solution of nitric acid except for the top plate | | good data =699+674+632+624+616-609 |