PMT counts over fixed time period
PMT 14 was biased to -1100V and a dual scaler (counter) NIM module is allowed to collect counts for 5 minutes. Five runs are taken for each arrangement (see table 1) of the source, scintillator, and collimator.
| Table 1: experimental configuration | ||||
|---|---|---|---|---|
| Config. # | Source | Scintillator | Collimator | Shielding |
| 1 | None | 1-66-x | None | None |
| 2 | Am-241 | 1-66-x | None | None |
| 3 | None | 1-66-x | Copper | None |
| 4 | Am-241 | 1-66-x | Copper | None |
| 5 | None | 1-66-x | Copper | Lead |
| 6 | Am-241 | 1-66-x | Copper | Lead |
| 7 | None | None | None | None |
| 8 | Am-241 | None | None | None |
With Scintillator
| Table 2: Counts/(5 min) with Scintillator | ||
|---|---|---|
| Run # | Am-241 | No Source |
| 1 | 2220 | 1743 |
| 2 | 2224 | 1733 |
| 3 | 2206 | 1610 |
| 4 | 2133 | 1692 |
| 5 | 2173 | 1703 |
| Average | 2191 | 1696 |
| Error of Avg. | 17 | 23 |
We observe a statistically significant difference in the number of counts between Am-241 and no source when the scintillator is attached to the PMT.
With Collimator, Scintillator
| Table 3: Counts/(5 min) with Collimator & Scintillator | ||
|---|---|---|
| Run # | Am-241 | No Source |
| 1 | 1738 | 1685 |
| 2 | 1740 | 1727 |
| 3 | 1724 | 1700 |
| 4 | 1786 | 1688 |
| 5 | 1715 | 1691 |
| Average | 1741 | 1698 |
| Error of Avg. | 12 | 8 |
There is a significant difference between the Am-241 counts and background counts, but the effect is much less pronounced than with no collimator.
With Collimator, Scintillator, Shielding
For these runs, the same setup as used above is surrounded with lead bricks to shield the scintillator from cosmic rays.
| Table 4: Counts/(5 min) with Collimator & Scintillator | ||
|---|---|---|
| Run # | Am-241 | No Source |
| 1 | 497 | 498 |
| 2 | 543 | 529 |
| 3 | 529 | 525 |
| 4 | 515 | 502 |
| 5 | 523 | 474 |
| Average | 521 | 506 |
| Error of Avg. | 8 | 10 |
No Scintillator
| Table 5: Counts/(5 min) with no Scintillator | ||
|---|---|---|
| Run # | Am-241 | No Source |
| 1 | 0 | 0 |
| 2 | 2 | 3 |
| 3 | 3 | 1 |
| 4 | 2 | 2 |
| 5 | 2 | 2 |
| Average | 1.8 | 1.6 |
| Error of Avg. | 0.4 | 0.5 |
We observe no significant difference between Am-241 and no source when no scintillator is used.