UMN CDMS Wiki archive
start › cdms › plasticscintillator › ebook › 20171116_nai

NaI Detector Characterization

Introduction

A Na-22 gamma source was placed underneath the NaI detector. Biases of +1000V and +1100V were applied to the Ortec 266 PMT (make sure it's a positive bias!!). The signals of the anode (negative pulses) and dynode (positive pulses) were examined.

Data

Detector B

Figure 2: Pulse Area Figure 3: Pulse Amplitude
Figure 4: Pulse Area Figure 5: Pulse Amplitude
Figure 6: Pulse Area Figure 7: Pulse Amplitude
Figure 8: Pulse Area Figure 9: Pulse Amplitude

Detector C

Figure 10: Pulse Area Figure 11: Pulse Amplitude
Figure 12: Pulse Area Figure 13: Pulse Amplitude
Figure 14: Pulse Area Figure 15: Pulse Amplitude
Figure 16: Pulse Area Figure 17: Pulse Amplitude

Analysis

Table 1: Summary of Results (Area Analysis)
Detector Bias (V) Output 511 keV Peak (nVs) 1.274 MeV Peak (nVs) Full Width at Half Max (nsec)
B 1000 dynode 3.06 7.69 248
C 1000 dynode 3.99 9.90 235
B 1100 dynode 6.48 15.36 243
C 1100 dynode 7.96 18.39 236
B 1000 anode 5.63 14.4 270
C 1000 anode 6.33 15.91 256
B 1100 anode 10.90 27.29 252
C 1100 anode 12.26 29.38 238