DCRC v2 test

Charge channel test

Phonon channel test

After fixing the readout vi, I managed to get the modulation curves on all 4 channels. The modulation depths are roughly same as DCRC v1: A and D are about 200mV, and B and C are about 450mV. After that I could not lock the squid nor do any further measurements. The board was sent to Fermilab on May 10 to do a further check.

DCRC v1

Superconducting resistance by IbIs

The table below can also be found in the run diary

Senser RSC (mΩ) RN (Ω)
A 1.6 1.4
B 2.5 1.8
C
D 2.3 1.8

Results from Run 2 can be found here.

The superconducting resistance is considerably lower than the results from CWRU or UCB. Vuk suggested the discrepancy may be because we didn't know the exact phonon channel gain.

Ba-133 pulse data

6 hours of Ba-133 data was taken after baking the detector for 0.5 hour. I posted a few plots below.

  • Charge overall calibration and crosstalk correction is done, position correction is not.
  • The charge spectrum has a lower energy resolution and is different from Soudan .
  • Loss of neutralization happened at 1.5e4 sec (4 hours) ?
Cut: No cut
Cut: No cut
Cut: Pt > 5 & inrange(Xdel, -50, 50) & inrange(Ydel, -50, 50)
Cut: 0.93 < Qpart & Qpart < 1.07 & 5 < Qsum & Qsum < 400
Cut: Pt>5 & inrange(Xdel, -50, 50) & inrange(Ydel, -50, 50) & cQchi2
Cut: Pt>5 & inrange(Xdel, -50, 50) & inrange(Ydel, -50, 50) & cQchi2
Cut: No cut
Cut: 5 < Qsum & Qsum < 200 & cQchi2 & cQin
Cut: 5 < Qsum & Qsum < 200 & cQchi2 & cQin
Cut: cQin & cQchi2 & 5<Qsum & Qsum<100
Cut: cQin & cQchi2 & 5<Qsum & Qsum<100
Cut: cQin & cQchi2 & 5<Qsum & Qsum<100
Cut: cQin & cQchi2 & 5<Qsum & Qsum<100