Table of Contents
100mm crystal ionization test Run11, Run13
Two devices G101a and G102a have been tested so far. The crystals are 100mm diameter and 33.3 thick. Fig. A1 shows the ionization electrode design on one side of the crystal. A ground plane is deposited on the other side of the crystal.
Run 11, G101a
More complete run summary can be found at here and here.
- G101 had a scratch across Q2 to Q4 on the electrode side.
- Charge return wire was only bonded to Phonon A on DIB 1.
- 4 collimated Am-241 sources (~20 Hz each) were placed ~2mm above the 4 electrodes.
Charge spectrum
- 60 keV peaks observed on Q1, Q2, and Q4.
- Q1 and Q3 were much weaker than Q2 and Q4.
- Charge collection efficiency was calculated based on Q2.
Neutralization
- The detector could hold neutralization for ~0.4 hours at -4V bias.
- Neutralization time increased with voltage bias magnitude.
- This provides a way to study charge trapping in the crystal.
Shared events
Run 13 G102a
More complete summary can be found at here
- Charge return wires are bonded to Phonon B on DIB 1 and Phonon C and D on DIB 2.
- 4 collimated Am-241 sources (~20 Hz each) were placed ~2mm above the 4 electrodes.
Charge spectrum
- 60keV peak observed on all 4 channels.
- 4 electrodes were better balanced than G101a.
- Charge collection efficiency calculated for 3 inner channels.
Neutralization
- G102a could hold neutralization ~15min.
- Good neutralization time increased with bias.
Shared events
- Share events were still observed between Q1 and Q3, Q2 and Q4.
Run 22
Summary for SNOlab DTC R&D meeting
Low Energy Events
The justification for this run lays in the discrepancy found between R16 and R20's results with G102a. Please refer to Hassan's preliminary comparison between the two runs for details (http://www.hep.umn.edu/cdms/cdms_restricted/K100/runs/run020/compareRuns/k100Run16Run20ComparisonPrelim.pdf). Essentially, in R20 we found that there were a large abundance of low energy events which were not present in R16. Note that in R16 Q4 was not working.
The source configuration for this run is different from the previous because the source has been placed next to the opposite face of the detector. The first round of taking data included series with the detector biased at -4V, -6V, -8V, and -10V and exposed to the internal Am-241 source. In these data series, the spectra looked relatively normal (https://zzz.physics.umn.edu/cdms/doku.php?id=cdms:k100:run_diary:20111204). There were a couple of positive things to take from this: all four channels seemed to be working properly, and no low energy event peaks were found similar to those in R20. The chi^2 vs. charge plots for -8V and -10V, however, featured a double band. An investigation was made into this and will be mentioned a bit later.
Following the promising results from the first 4 series, the next 3 data series were taken with the detector biased at -4V, -8V, and -13V and exposed to an external Ba-133 source. Surprisingly, the spectra for these biases revealed significant low energy event peaks (https://zzz.physics.umn.edu/cdms/doku.php?id=cdms:k100:run_diary:20111205). This is interesting to note because in R20 the spectra for -13V bias exposed to Ba-133 seemed to make the low energy events disappear. It seems from the data taken in this run, though, that this disputes that. The low energy event peak seems to be source-independent.
It is important to note that after the first day of taking data, each data-taking sequence afterwards experienced an interruption with the DAQ crashing. The implications of this are significant. When the DAQ crashed overnight, it left the detector biased for hours. Usually we flash between data series to reverse the effects of degradation. In these cases, though, the detector is usually only biased for an hour at a time. Perhaps it would be necessary to flash longer due to the extensive bias time from the DAQ crashing. We flashed the detector overnight for a total of 10 hours of flashing. The detector was biased at -13V with the Am-241 source. The chi^2 vs. charge and charge spectra for this data set can be seen in figures B1 and B2. Below Fig. B1 and B2 are the plots from data with the detector taken under the same conditions except for the extensive flashing (and post-DAQ crash).
-13V, Am-241 Data - Long flashing (10 hrs)
| No low energy events as seen in the -4,-8,-13V bias set. Note that the 60 keV peak is quite clear in Q1 and Q2. |
|---|
-13V, Am-241 Data
| Above are the results from the data taken under the same conditions without the long flashing period. The 60 keV peak can only clearly be seen in Q1. |
|---|
Comparing the two -13V/Am-241 data sets, it makes a strong support for the argument that there is a lot of degradation to overcome in the detector when the DAQ crashed. The normal spectra is achievable, though, as seen in Fig's B1 and B2. This speaks well for the functioning of the detector.
Following in this investigation are long series of taking data to see the behavior of the charge spectra, etc, over time. The first set is a 12 hour run of taking data. The subsequent set after that is an ongoing data series that is being analyzed periodically to see if/when the low energy events appear.
-10V, Ba-133 Data, 12 hours of taking data (First 3 hours only)
-10V, Ba-133 Data, 12 hours of taking data (Last hour)
Here begins the continous long series, referred to as Long Series #2.
-10V, Ba-133 Data, Long Series #2 - Hour:1st-3rd (0-10800s)
-10V, Ba-133 Data, Hour: 12th-15th (43200-54000s)
-10V, Ba-133 Data, Hour: 21st-23rd (75600-82800s)
-10V, Ba-133 Data, Hour:29th-31st (104400-111600s)
-10V, Ba-133 Data, Hour:36th-39th (140400-129600s)
Double Band Feature
Recall from the -4V → -10V data set that a double band feature appeared in the chi^2 vs. charge plot in Q3 for -8V and -10V. In all subsequent data sets with the large low energy event peak, a significant spread in the chi^2 vs. charge is present. Below are some quick investigations into the nature of the bands/spread.
In Figs. C1a and C2a, we make a distinction between the 3 series of data taken at each bias. Clearly, the spread is occurring because the chi^2 values between the series behave differently. It is not in a consistent manner, though, between different biases. For -8V the third series has much higher chi^2 values than the first two series, while in the -10V set its chi^2 values are lower. We looked at the PSD's for these biases to see if there was a clear correlation between chi^2 and the PSD, and for Q3 we do see a slightly wild behavior in the PSD.
PSD's
We decided to also compare the PSD's for the -13V sets presented earlier (Figs. B1-B4) for their low event characteristics.
It's interesting to see that in the long flashing spectra, the first series has only one peak in the beginning. One thought is that having double peaks at the beginning may be due to the switching on/off of lights in the lab.
Run 22 Quality Data Summary
Below is a list of the acceptable and bad data sets at various biases. There were several biases that had multiple retakes, so if at one bias, say, -13V with Ba-133, had good and bad data, I've only listed the good data below. However, if a bias has only bad data, say +3V with Ba-133, then I've listed it below.
| Acceptable Data Sets | Series | Comments | Bad Data Sets | Series | Comments |
|---|---|---|---|---|---|
| +2V Ba | 07111221_0019 07111221_0139 07111221_0300 | Neutralization barely lasts; Q1 spectra is ok- rest are not as good; charge baseline varies | |||
| -2V Ba | 07111215_2149 07111215_2311 07111216_0031 | ||||
| -3V Ba | 07111220_2014 07111220_2135 07111220_2255 | Neutralization doesn't last long; baseline varies | +3V Ba | 07111221_0424 07111221_0545 | Large spread in chi2; Q1 spectra is okay, others are weird |
| -4V Ba | 07111219_1527 07111219_1647 07111219_1809 | +4V Ba | 07111206_2115 07111206_2235 | Spread in Q2,4 chi2; low E events | |
| -6V Ba | 07111216_0156 07111216_0316 07111216_0437 | ||||
| -8V Ba | 07111219_1932 07111219_2052 07111219_2213 | baseline kind of jumpy; second series doesn't seem to start off at a higher neutralization settingas the first and third | +13V Am | 07111205_1456 07111205_1616 07111205_1736 | Low E events; multi-band structure in Q2,4 chi2 |
| -10V Ba | 07111209_2009 07111210_1937 07111216_0602 07111216_0722 07111216_0843 | First set is 12 hours of data taking; Next set is the 2 days of data taking | |||
| -13V Ba | 07111219_2338 07111220_0058 07111220_0218 | ||||
| +13V Ba | 07111220_0343 07111220_0503 07111220_0624 | Multi-bands in chi2, but distinguishable; baseline is a bit crazy; spectra look ok | |||
| -3V Am | 07111224_0444 07111224_0604 07111224_0725 | Q3, Q4 are not very good; chi2 is fine | |||
| -4V Am | 07111203_18 07111203_19 07111203_20 | ||||
| -6V Am | 07111203_22 07111203_23 07111204_00 | ||||
| -8V Am | 07111204_021 07111204_033 07111204_045 | Double band in Q3 chi2 - note discussion above with PSD's | |||
| +8V Am | 07111223_1628 07111223_1748 07111223_1908 | Two charge channels are okay, chi2 are fine | |||
| -10V Am | 07111204_061 07111204_073 07111204_085 | Double band in Q3 chi2 | |||
| +10V Am | 07111223_2033 07111223_2153 07111223_2314 | Q3, Q4 are not good; chi2 is fine | |||
| +13V Am | 07111224_0039 07111224_0159 07111224_0319 | Q3, Q4 are not good; chi2 is fine | |||
| -13V Am | 07111208_1023 07111208_1144 | chi2 a bit shaky |




















































