2025/07/28 (Monday)

Previous Day

  • Lab in-personnel: Himangshu, Yan
0900 The lab desktop has been shutdown. Restarting temperature monitoring programs. The temperature is still stable at 77K.
0920 The LN tank is empty. Need to refill. and need to refill the cold trap soon!
0940 Close V8, V1, turn on the 3He dump, and open V3, V5, V2 to pump the mixture back to the dump. (condenser speedivalve and still gatevalve is open.)
0955 Prepare to pump out IVC. First turn on scroll, open the scroll valve to pump out the IVC, then switch to turbo after 5 min. After switching to turbo, pressure gauge is reading 2e-4 mbar.
1030 Close V3, 5, 2, 9. Turn off He3 pump.
1035 turbo pressure gauge is reading 2e-5 mbar.
1250 MC temperature slowly increased from 77.24 K to 77.29 K. So we expect 0.05K/2 hours, so it will be 12*0.05=~0.6 K tomorrow noon time.
1440 Refill LN2 dewar from Refill Room Main Line
1503 Leak Detector on IVC = 1.5e-7 mbar L/s
17xx Starting electric check.
Detector: S104 Stripline 48
SC: none 77 K
Face: 3 Date 07/28/2025
SQUET: ZC-39 Cable 1
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 1.631 k OK
SQUID bias B/Ret 19 34 1.504 k OK
SQUID bias C/Ret 8 24 — k OK
SQUID bias D/Ret 23 39 — k OK
Feedback A/Ret 37 21 15 - 25 k 8.85 k OK
Feedback B/Ret 2 18 8.28 k OK
Feedback C/Ret 41 25 11.31 k OK
Feedback D/Ret 22 6 6.31 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 364, 310, 305, 304, 326 OK
QET B/Ret 35 29, 33, 47, 48 71, 64, 122, 55 OK
QET C/Ret 7 5, 9, 10, 28 77, 36, 27, 36 10Mohm
QET D/Ret 38 36, 40, 43, 44 31, 32, 31, 31 10 Mohm
QET A/QET B 3 35 1000-3000 open
QET A/QET C 3 7 open
QET A/QET D 3 38 open
QET B/QET C 35 7 open
QET B/QET D 35 38 open
QET C/QET D 7 38 6.1k
1557 Electrical check of ZC-70+interposer
Cooldown checkout sheet (rev C SQUET)
Detector: S104 Stripline 41
SC: none 77 K
Face: 5 Date 07/28/2025
SQUET: ZC-70 Cable 2
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k open OK
SQUID bias B/Ret 19 34 1.672 k 13 nS
SQUID bias C/Ret 8 24 1.763 k 11 Mohm, Squid and Feedback is shorted by 9 Mohm
SQUID bias D/Ret 23 39 1.996 k OK
Feedback A/Ret 37 21 15 - 25 k 2 Mohm OK
Feedback B/Ret 2 18 9.19 k Ok
Feedback C/Ret 41 25 13.83 k 28 nS
Feedback D/Ret 22 6 16.91 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 OL, 59.4, OL, OL, OL 11 Mohm
QET B/Ret 35 29, 33, 47, 48 64.8, 65.1, OL, OL 11 Mohm
QET C/Ret 7 5, 9, 10, 28 85.5, 48.5, OL, OL 11 Mohm
QET D/Ret 38 36, 40, 43, 44 43.8, 44.5, OL, 4 Mohm Ok
QET A/QET B 3 35 1000-3000 22.5 kohm ???
QET A/QET C 3 7 16.7 kOhm
QET A/QET D 3 38 open
QET B/QET C 35 7 6 kOhm ???
QET B/QET D 35 38 open ???
QET C/QET D 7 38 open
LED 1 chassis 42 1.0 - 1.1 V 1.38 1.105
LED 2 chassis 26 1.375 1.69
Motor Coil Y1 pBr3 pBr4 129
Motor Coil Y2 pCr3 pCr4 85.6
Motor Coil X1 pAr3 pAr4 80.6
Motor Coil X2 pAr5 pDr3 68.8
Xendstop pAr1 chassis open
Yendstop pDr4 chassis 1.3 Mohm
1931 3rd Electrical check of ZC-13 after connecting to Stripline 43 on Tower Face 2
Cooldown checkout sheet (rev C SQUET)
Detector: S104 Stripline 43
SC: none 77 K
Face: 2 Date 07/28/2025
SQUET: ZC-13 Cable 3
Signal Pos. lead Gnd lead Nominal Actual Chassis
(Gnd lead)
Nominal:
< 0.1 nS
SQUID bias A/Ret 4 20 3 - 5 k 1.734 k OK
SQUID bias B/Ret 19 34 1.854 k OK
SQUID bias C/Ret 8 24 — k OK
SQUID bias D/Ret 23 39 1.904 k OK
Feedback A/Ret 37 21 15 - 25 k 7.46 k OK
Feedback B/Ret 2 18 8.0 k OK
Feedback C/Ret 41 25 — k OK
Feedback D/Ret 22 6 8.41 k OK
QET A/Ret 3 1, 13, 14, 17, 32 60 - 80 106.1, 45.6, 41.3, 41.5, 42.8 OK
QET B/Ret 35 29, 33, 47, 48 91.6, 73.1, 115.6, 69.8 OK
QET C/Ret 7 5, 9, 10, 28 66.3, 41.6, 30.6, 39.1 11 Mohm
QET D/Ret 38 36, 40, 43, 44 64.2, 39.9, 38.9, 39.2 OK
QET A/QET B 3 35 1000-3000 open
QET A/QET C 3 7 open
QET A/QET D 3 38 open
QET B/QET C 35 7 open
QET B/QET D 35 38 open
QET C/QET D 7 38 open
Throughput test procedure, Cold Trap 1:
  1. All mixture in dump, LN2 trap 1 is cold, condenser & still valves open (3He pump on - optional)
  2. Open V9, V13a, V12a+speedivalve, V3. Wait for G1 to be close to G2 (within 100 mbar).
  3. Close V12a+speedivalve and V3, open V1
  4. monitor G1
1600 Perform mixture throughput test at MC = 77 K using above procedure with 3He pump on. Let G1 rise to 568 mbar before Step 3, then it falls to ~60 mbar, then rises. Throughput = 43.47 mbar/min.
1620 Start another round of cryocleaning.
1905 Collect mix back into the dump and let cold trap slowly warm to room temperature.