Assuming pumped out IVC and CVJ with turbo.
| IVC under turbo = 6.3e-5 mbar. | |
| Pump out fridge lines: |
- Make sure mixture is isolated in the dump.
- Check fridge pressures by opening condenser speedivalve+V1 and still gate valve: LHe trap+Condenser(G1)/Still(P1) = no change/1700 mTorr.
- Connect scroll pump to He-4 pump line, turn it on.
- Open pump line valve, V5A, V2A, V4, wait a minute, then open V5.
- You're now pumping on the fridge lines. Wait until P1 and G1 flatten out.
- Close condenser speedivalve and still gate valve, V1, V5, V2A, V5A and close and turn off the scroll pump.
| 1530 | G1 remain at zero level. P1 stable at 1700 mTorr |
|---|---|
| 1535 | Throughput test procedure, Cold Trap 1: |
- All mixture in dump, LN2 trap 1 cold, condenser & still valves open (3He pump on - optional)
- Open V9, V8+speedivalves, V3 and slowly open V5. Wait for G1 to be rise and then fall, close to G2.
- Close V3 and V5, open V1
- monitor G1
| 1245 | Perform mixture throughput test at MC = 300 K using above procedure with 3He pump on. Let G1 rise to 600 mbar before Step 3, then it falls to ~25 mbar on step 4 and starts to rise. Reaches ~50mbar after 6.5min Throughput = 3.8 mbar/min. (very rough estimation) |
|---|---|
| 1317 | Pump mixture back to dump: |
- Close V8 then turn on the 3He pump.
- Open V3, V5.
- Open V2 initially to marked 1, then further in steps to 30 degree, making sure G2 doesn't rise too quickly.
- Continue pumping until G1 stabilizes at a minimum value (typically a few mbar) and P1 is less than 100 mTorr.
- Stop pumping on cold trap 1, close coldtrap 1 speedivalves and V3.
- Stop pumping on fridge, close still gate valve, condenser speedivalve, V1, V5.
- Close V2 completely, close V9.
| 1207 | Placed cold trap 1 in LN2 bath |
|---|
| —- | Pump out 1kpot and main bath: |
|---|
- Get to baseline on G3 by opening scroll pump valve and V4A. G3 at 0 mbar.
- Close scroll pump valve and V4A
- Check pressure in 1K Pot on G3 by opening the 1kpot speedivalve. Open V4A and slowly pump out with Scroll Pump by opening the scroll pump valve slowly connected to the 4He line.
- Close the 1kpot speedivalve and V4A. close the scroll pump valve.
- Open V1A which opens up the line to the main bath (this will cause the line to expand as the gas fills it).
- Open V4A and pump on Main Bath with scroll pump, making sure not to overload the scroll pump.
- Once P2 < 2 mbar,close V1A and open 1K Pot speedivalve and needle valve (5 turns) to pump out 1K pot volume to < 10e-1 mbar on P2.
| 1214 | Check pressure in 1K Pot by opening the 1kpot speedivalve. G3 went up, but below the 1st least count mark ~ 50 mbar. Pump out with Scroll Pump |
|---|---|
| 1218 | Close 1kpot speedivalve, close scroll pump valve, open V1A which opens up the line to the main bath (this will cause the line to expand as the gas fills it). G3 reads 950 mbar |
| 1222 | Open V4A and pump on Main Bath with scroll pump, making sure not to overload the scroll pump. |
| 1230 | P2 < 2 mbar, close V1A. Open 1K Pot speedivalve and needle valve (5 turns) to pump out 1K pot volume to < 10e-1 mbar on P2. |
| 1252 | Once P2< 10e-1 mbar. Close V4A and scroll pump valve. |
|---|
| Flush the 1kpot and main bath: |
- With the scroll pump pumping on the He-4 pump line, open V5A, V2A, and the vent speedivalve to evacuate the clear hose leading to the He gas line.
- Once P2 falls to <10^-1 mbar, close the He-4 pump line valve and turn off the He-4 pump or scroll pump.
- Open V1A, V4A, V5A, V2A.
- Open the blue valve that connects gas line to IGH - gas is filling the main bath now.
- Once the mechanical gauge (connected to the 1kpot speedivalve) reaches 15mmHg, the bath is at ~0.5bar, stop flushing. Close V5A.
- Pump out with the scroll pump to < 4 mbar on P2, making sure not to overload the pump.
- Close the scroll pump valve and repeat steps 3 to 5 once more
- Close the scroll pump and pressurize the main bath and 1K Pot with helium to a few 100 mbar+1atm on G3, then close 1K Pot speedivalve and needle valve.
- Close V1A, V4A, V5A, V2A.
| 1244 | Close off scroll pump, then pressurize bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar on P2. |
|---|---|
| 1251 | Close off scroll pump, then pressurize bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar on P2. |
| 1258 | Pressurize main bath and 1K Pot (both sides) with helium to 1000 mb on G3, then close 1K Pot speedivalve and needle valve, V1a, V4a, V5a, V2a. |
| We won't flow mixture through the fridge as I would like to do another cryocleaning today | |
| – | We will flow mixture through the fridge during LN2 cooldown. Let mixture from dump through cold LN2 Trap 1 to fridge, both sides. Then isolate the sides and circulate with V6 open partially and mix open to dump. V9, V13a, V12a+speedi, V1, Condenser speedi, Still gate valves open, 3He pump on. Open V6 to 12.6%. |
| – | Flow mixture through fridge: |
- Open V9, V13a, V12a+speedivalve, V1, condenser speedivalve, still gate, V3, let things settle.
- Turn on 3He pump, after a min, open V6 to 12.6% and close V3.
| – | P1(Condenser) → 0.05 mbar. (forgot to close V3 right away) |
|---|---|
| – | Turbo on IVC = 12nbar |
| – | Try closing V12a+speedivalve. G1 falls as it should, showing we have throughput. Falls at 10 mbar/min . Open V12a+speedivalve again. |
| 1408 | Let 6.3 mbar He exchange gas into IVC, this is a little less than one hose length (connected to the mechanical gauge on the turbo pump) |
|---|---|
| 1410 | MC(Cernox) = 296.6 K, Sorb = >249 K (max), IVC = 6.3 mbar, G2/G1/P1 (circulation) = –/–/– mbar, Level resistor = 0.990 kΩ. |
| 1450 | Inserted LN2 transfer stick, connected to the LN2 dewar with a hose, inserted the level resistor, disconnected the helium recovery line. |
|---|
| 1420 | Begin LN2 transfer |
|---|---|
| 1430 | MC(Cernox) = 292.7 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1/P1 (circulation) =– mTorr, Level resistor = 1.111 kΩ. |
| 1445 | MC(Cernox) = 272 K, Sorb = >249 K (max), IVC = 4.0 mbar, G2/G1/P1 (circulation) =— mTorr, Level resistor = 1.290 kΩ. |
| 1508 | MC(Cernox) = 199 K, IVC = 2.5 mbar, G2/G1/P1 (circulation) = — mTorr, Level resistor = 1.354 kΩ. |
| 1531 | MC(Cernox) = 158.2 K, IVC = 2.2 mbar, G2/G1/P1 (circulation) =— mTorr, Level resistor = 1.354 kΩ. |
| 1539 | MC(Cernox) = 150.0 K, IVC = 2.2 mbar, G2/G1/P1 (circulation) =— mTorr, Level resistor = 1.354 kΩ |
| Level resister is stable, we know we dont need the full LN2 dewar, end LN2 transfer | |
| 1540 | |
| 1542 | Stop LN2 transfer. Close LN2 Dewar valve. Pull out Level Resistor and put stopper in port. Put ping-pong ball on MB exhaust port. |
| 1555 | P1 < 100 mTorr. Close V3+speedivalves, V5, V2 then V9 and turn off 3He pump. |