| Status | Bagging all valves near 1K pot as we found multiple valves leaky |
| 1005 | Turbo connections redone. Checked pressure stability with time in the bellow, more than 10 minutes under 10e-3 mbar |
| 1210 | Main Bath pumped out with LD, Needle valve closed, No signal |
| | |
| 1300 | Bagged all three valves: 1K pot Speedivalve, 1K pot relief valve, 1K pot needle valve with ziplocks. Sprayed He around the bags, avoiding lower areas to see false signal. LD on 1K: 5e-7 mbar L/s (bg: 7.5e-8 mbar L/s) |
| 1310 | Sprayed inside speedivalve bag: signal: 2e-7 mbar L/s. Still smaller than outside spray, suggesting it is away from the leakest valve |
| 1325 | Sprayed inside 1k pot needle valve: signal 8.5e-7 mbar L/s, bigger than outside spray |
| | |
| 1400 | Sprayed inside relief valve bag: signal 2.85e-4 mbar L/s. Very big |
| | |