* Lab in-personnel: Himangshu

Check Leak detector level today.
It was above 10e-0 since we added exchange gas to warm up, but temp of MC still at 26.5 K
Pump out the IVC to prepare for leak checking from Main Bath
Leak detector baseline after turbo pumping: 2.1e-6 mbar L/s
Let He gas into Main Bath through LN2 transfer stick on top of the fridge
No change in Leak detector while I stream through 4 psi He gas into the pipe!
CVJ at 0.54 mbar. Not adding anymore He gas in this one