2025/01/29 (Wednesday)

Previous Day

* Lab in-personnel: Himangshu

1243 Connected Leak detector to IVC. With turbo on IVC line (not yet open to IVC volume), Leak = 3.5 pbar L/s
1245 With turbo valved out, IVC line (not yet open to IVC volume), Leak = 7.4 pbar L/s
1250 With turbo valved out, Leak detector on IVC, Leak= 4e-2 mbar L/s , IVC pressure: 10 ubar
1255 Pump out IVC with turbo
1256 IVC: 4e-4 mbar , Leak: 8.7e-3 mbar 1/s
1258 IVC: 2.5e-4 mbar, leak: 6.7e-3 mbar 1/s
1300 IVC: 1.9e-4 mbar, leak: 4.8e-3 mbar 1/s
1303 IVC: 1.6e-4 mbar, leak: 3.5e-3 mbar 1/s
1604 Stopped Pumping on the IVC. Leak detector: 1.3e-4 mbar 1/s. Close IVC ports on top of fridge to see numbers tomorrow morning
Check leak detector next day, Leak detector: 1.6e-4 mbar 1/s. Looks like a small leak.