| 1120 | Baratron: 5.259 V (== 5.259*1000/7.6 = 691.9 mbar), G2 = 690 mbar |
| 1130 | Fill LN2 cold trap bath from bomb dewar. |
| 1230 | MC(H35)/Sorb/1K Pot = 4.63 K/–/– K, G1/G2 = 460/450 mbar, MB = 95.00%, P1 = offscale (analogue), 350 Torr (digital) |
| | (Closed)Close IVC valves on top of fridge. Mixture already in fridge ciculation loop |
| 1240 | Pump out line leading to 1K Pot with scroll pump (V4a open, V1a closed, 1kpot speedivalve closed). Then switch the line over to the 4He pump. Turn on 4He pump and make sure the pump exhaust is connected to recovery and open valve to recovery lines. |
| 1141 | Check pressure in 1K Pot by opening the 1kpot speedivalve with V4a closed: Mechanical gauge = ~360 mbar on G3. Baratron: 3.502 V |
| 1241 | Pump out 1K Pot by slowly opening V4a then the manual valve at the 4He pump until it just starts to make noise, once it quiets down somewhat open it up a little more. continue this until the valve is fully open. |
| | Leaving mixture in the fridge when starting 1K pot. P2 < 1 mbar. 1K Pot = – K(– ohm) |
| 1150 | Open 1K Pot needle valve → 30° (half the unit on the needle marks). 1K Pot → - / 20 mbar, Baratron: 3.405 V |
| 1153 | Open 1K Pot needle valve → 45°. 1K Pot → – K / 60 mbar. Closing back to 30° |
| 1154 | Open 1K Pot needle valve → 30°. 1K Pot → – K (– ohm) / 25 mbar. Baratron: 2.900 V, MC(H35) = 3.23 K |
| 1155 | Temps start to drop, mixture pressure starts to drop |
| 1157 | MC(H35) temp drops, linear at around 3.11 K, Baratron: 2.535 V |
| 1204 | MC(H35) = 2.80, 1K Pot = – / 25 mbar. G1 (fridge+dump) = 240 mbar, Baratron: 1.74 V. |
| 1229 | MC(H35) = 2.04, 1K Pot = –(– ohm) / 25 mbar. G1 (fridge+dump) = 50 mbar, Baratron: 0.454 V. |
| 1232 | We will use up all mixture this time. Close V5 and V9, leaving – mbar in the dump. Turn on 3He pump. |
| 1249 | Baratron: 0.395. Start circulation through Trap 1. V2 → Mark 1+ mark2. Condenser(G1) peaks at 350 mbar. P2 jumps to 100 mbar before falling again to 20 mbar. MC(Cernox) → 1.53 K. Still(P1) → - K / 10 Torr, Coldplate = – K, 1K Pot = – |
| 1301 | V2 → mark 3. Condenser(G1) peaks at 450 mbar. MC(H35) → 1.36 K. Still(P1) →– K / 5 Torr, Coldplate = – K, 1K Pot = –, Baratron: 0.390 V |
| 1310 | G1/G2 keeps growing over time above 600 mbar, so i opened the dump valve 9 to lower it back to 450 mbar range. Baratron: 0.589 V |