2024/12/18 (Wednesday)

Previous Day

  • Lab in-personnel: Himangshu
Status Run 82 Stable Base at 45 mK.
MC(H35) = 45 mK, Tower(U02844) = 112 mK, flat.
1230 Turned on DCRCs
SQUID Tuning: DCRC3 on Cable 1
DCRC3pA: Bias: -58.8, Lockpoint: -3.73
DCRC3pB: Bias: -85.1, Lockpoint: -3.42
DCRC3pD: Bias: -75.3, Lockpoint: -3.62
SQUID Tuning: DCRC3 on Cable 1
DCRC3pA: Bias: -58.8, Lockpoint: -3.73
DCRC3pB: Bias: -85.1, Lockpoint: -3.42
SQUID Tuning: DCRC3 on Cable 2
DCRC3pD: Bias: -80.6, Lockpoint: -2.30
Added mixture chamber heater wiring. R_MC = 475 ohm, R_Still = 469 ohm . The maximum current that can be sourced from sample heater output is 100 mA. Max Power = 1W, With R_MC = 475 ohm, P = Ie+2 R ⇒ 1 W = (0.1)**2 R ⇒ R = 100 ohm. Maximum power for a range is the lower of the two; P = I2R and P = V2/R where P is heater power in watts, I is the maximum current for the range and V is the compliance voltage of 10 V. For R = 475 ohm, I(100mA) P = 210 mW, I(31.6 mA) P = 474 210 mW
Adding 50 uW of power to MC.