2024/11/22 (Friday)
- Lab in-personnel: Himangshu
| Status | Run 82 In Preparation. |
|---|---|
| LN2 transfer yesterday. Cernox stuck at 100K overnight. Most probable cause is we ran out of exchange gas. I reintroduced 4 mbar exchange gas after pumping out the IVC and added more LN2. Temp started to drop further | |
| Leak detector shows a reading in 10e-6 mbarL/s level which is very high to check leaks. So, we decided to refill LN2 and do the blowout on Monday, turbo pumping the IVC over the weekend |