2024/01/31 (Wednesday)
- Lab in-personnel: Himangshu
| Status | Run 79 beginning. Fridge at base temperature. |
|---|---|
| 1220 | Refilled LN2 Cold Trap 1 from dewar. |
| 1200 | Tower/MC(H35)/Coldplate = 66.5/34.9/147 mK, Still = 1.11 K / 0.110 mbar / 0 mW, G1/G2= 69/76 mbar, 1K Pot = 45° / 1.64 K / 24 mbar. MB = 33.5% . No DCRCS are turned ON yet for today. |
| 1232 | Start MIDAS in cdms2(if not already running). Power ON the DCRC3 (donot plug the IO cable yet) using the ethernet cable. Start Frontend GUIs(dcrc_driver01, towerfe3_01, triggerfe2). (Open SQUIDTuningGUI, select correct DCRC and click SQUID Control, Select correct DCRC and Initialize each phonon channel if not done. Close SQUID Control) , Stop Frontend, Unplug ethernet cable, connect IO cable, Start Frontend again |
|---|
| 1247 | DCRC3 is connected to Cable2+diverter to get squid tuning parameters |
|---|---|
| Cable 2: SQuid Bias: 0, -85, -81.36, -83.1 LockPoint: 0, -4.98, -4.42, -4.11 | |
| 1257 | Check response to QET bias changes. pB,pC,pD show response but no pulses yet. Tower temp went upwards to 71 mK from 62 mK after connecting this cable |
| 1305 | Stop Run(if run in progress). click SQUID Control, Initialize all phonon channels. Close SQUID Control. Stop Frontends. Unplug ethernet cable to power off the DCRC |
| 1316 | DCRC3 is initialized. Connect Cable 1 and Start frontends(dcrc_driver01, towerfe3_01, triggerfe2). Set SQUID bias and lockpoints. (SQuid Bias: -89.8, -91.6, -74.6, -100.5 LockPoint: -8.28, -5.07, -4.21, -7.36 ). Need to change Open Loop to Non-inverting and Pre-Amp to Feedback as well |
| Tower temp went from 62.9 mK to 63 and came back down to 62.7 mK | |
| 1337 | Zap SQUID since pattern on pA is not making sense |
| 1420 | Need to change Open Loop to Non-inverting and Pre-Amp to Feedback as well |
| 1540 | DCRC3 is initialized. Connect Cable 1 and Start frontends(dcrc_driver01, towerfe3_01, triggerfe2). Set SQUID bias and lockpoints. (SQuid Bias: -91.7, -88.4, -74.6, -100.5 LockPoint: -9.45, -3.01, -4.21, -7.36 ). Need to change Open Loop to Non-inverting and Pre-Amp to Feedback as well |
|---|
| 1545 | DCRC3 is initialized. Connect Cable 2+diverter and Start frontends(dcrc_driver01, towerfe3_01, triggerfe2). Set SQUID bias and lockpoints. (SQuid Bias: 0, -85, -81.36, -83.1 LockPoint: 0, -4.98, -4.42, -4.11 ). Need to change Open Loop to Non-inverting and Pre-Amp to Feedback as well |
|---|---|
| 1617 | No pulses seen on Cable 2. Tower temp rises from 62.8 mK to 70.6 mK since this DCRC is on. |
| 1641 | DCRC3 is initialized. Connect Cable 3 and Start frontends(dcrc_driver01, towerfe3_01, triggerfe2). Set SQUID bias and lockpoints. (SQuid Bias: -70.4, -101.7, 0, -75.4 LockPoint: -4.55, -4.65, 0, -2.64 ). Need to change Open Loop to Non-inverting and Pre-Amp to Feedback as well |
|---|