2024/01/29 (Monday)
- Lab in-personnel: Himangshu, Ping
| 1155 | Refilled LN2 Cold Trap 1 from bomb dewar. Partially filled, need to figure out how to unscrew the LN2 dewar plug |
|---|---|
| 1400 | Pump out line leading to 1K Pot with scroll pump (V4a open, 1kpot speedivalve closed). Then switch the line over to the 4He pump. Turn on 4He pump and make sure the pump exhaust is connected to recovery and open valve to recovery lines. |
| 1410 | Check pressure in 1K Pot by opening the 1kpot speedivalve with V4a closed: Mechanical gauge = 0.33 bar. |
| 1324 | Pump out 1K Pot by slowly opening V4a then the manual valve at the 4He pump until it just starts to make noise, once it quiets down somewhat open it up a little more. continue this until the valve is fully open. |
| 1330 | MC(Cernox)/Sorb/MC(H35) = 4.37/4.0/4.35, 1K Pot = 3.13 K / 0.95 mbar. G1 (fridge+dump) = 413 mbar. MB = 57.9% |
| Detector: | S104 | Stripline | 41 | ||
| SC: | none | 77 K | |||
|---|---|---|---|---|---|
| Face: | 5 | Date | 01/24/2024 | ||
| SQUET: | ZC-70 | Cable | 2 | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | – k | OK |
| SQUID bias B/Ret | 19 | 34 | 0.088 k | OK | |
| SQUID bias C/Ret | 8 | 24 | 0.090 k | OK | |
| SQUID bias D/Ret | 23 | 39 | 0.100 k | OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 1.236k k | OK |
| Feedback B/Ret | 2 | 18 | 1.226 k | OK | |
| Feedback C/Ret | 41 | 25 | 1.232 k | OK | |
| Feedback D/Ret | 22 | 6 | 1.217 k | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | OL, 73.5 ,OL, OL, OL | NOT OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 46.3, 45.9, OL, OL | OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 41.7, 32.7, OL, OL | OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 38.3, 39.0, OL, OL | OK | |
| QET A/QET B | 3 | 35 | 1000-3000 | open | |
| QET A/QET C | 3 | 7 | open | ||
| QET A/QET D | 3 | 38 | open | ||
| QET B/QET C | 35 | 7 | 0.088K | ||
| QET B/QET D | 35 | 38 | open | ||
| QET C/QET D | 7 | 38 | open | ||
| open | |||||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | open | |
| LED 2 | chassis | 26 | 3.69 k | ||
| Motor Coil X1 | 169 | OPEN | |||
| Motor Coil X2 | 169 | 0.036 k | |||
| Motor Coil Y1 | 169 | 0.034 k | |||
| Motor Coil Y2 | 169 | 0.02 k | |||
| Xendstop | 1k | - | |||
| Yendstop | 1k | 0.028 k | |||
| 1355 | Electrical check of ZC-13 after connecting to Stripline 43 on Tower Face 2 on Cable 3 | ||||
|---|---|---|---|---|---|
| Detector: | S104 | Stripline | 43 | ||
| SC: | none | 77 K | |||
| Face: | 2 | Date | 01/24/2024 | ||
| SQUET: | ZC-13 | Cable | 3 | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 0.087 k | OK |
| SQUID bias B/Ret | 19 | 34 | 0.102 k | OK | |
| SQUID bias C/Ret | 8 | 24 | – k | OK | |
| SQUID bias D/Ret | 23 | 39 | 0.098 k | NOT OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 1.237 k | OK |
| Feedback B/Ret | 2 | 18 | 1.241 k | OK | |
| Feedback C/Ret | 41 | 25 | – k | OK | |
| Feedback D/Ret | 22 | 6 | 1.219 k | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 51.0, 31.6, 29.0, 29.2, 33.0 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 30.4, 30.6, 44.0, 30.4 | OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 34.6, 27.5, 18.3, 26.6 | OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 58.0, 47.2, 45.8, 45.7.8 | OK | |
| QET A/QET B | 3 | 35 | 1000-3000 | open | |
| QET A/QET C | 3 | 7 | open | ||
| QET A/QET D | 3 | 38 | open | ||
| QET B/QET C | 35 | 7 | open | ||
| QET B/QET D | 35 | 38 | open | ||
| QET C/QET D | 7 | 38 | open | ||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | - | |
| LED 2 | chassis | 26 | - | ||
| Motor Coil X1 | 169 | 0.025 k | |||
| Motor Coil X2 | 169 | open | |||
| Motor Coil Y1 | 169 | 0.041 k | |||
| Motor Coil Y2 | 169 | 0.017 k | |||
| Xendstop | 1k | - | |||
| Yendstop | 1k | - | |||
| 1356 | Electrical check of ZC-31 after connecting to Stripline 48 on Tower Face 3 on Cable 1 | ||||
|---|---|---|---|---|---|
| Detector: | S104 | Stripline | 48 | ||
| SC: | none | 77 K | |||
| Face: | 3 | Date | 01/24/2024 | ||
| SQUET: | ZC-31 | Cable | 1 | ||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis |
| (Gnd lead) | |||||
| Nominal: | |||||
| < 0.1 nS | |||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 0.106 k | OK |
| SQUID bias B/Ret | 19 | 34 | 0.075 k | OK | |
| SQUID bias C/Ret | 8 | 24 | 0.105 k | OK | |
| SQUID bias D/Ret | 23 | 39 | 0.095 k | OK | |
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 1.223 k | OK |
| Feedback B/Ret | 2 | 18 | 1.233 k | OK | |
| Feedback C/Ret | 41 | 25 | 1.229 k | OK | |
| Feedback D/Ret | 22 | 6 | 1.217 k | OK | |
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 72.1, 26.7, 23.0, 23.3, 29.5 | OK |
| QET B/Ret | 35 | 29, 33, 47, 48 | 42.7, 40.7, 43.1, 39.4 | OK | |
| QET C/Ret | 7 | 5, 9, 10, 28 | 30.2, 22.9, 15.2, 38.5 | OK | |
| QET D/Ret | 38 | 36, 40, 43, 44 | 237.8, 235.8, 235.3, 235.6 | OK | |
| QET A/QET B | 3 | 35 | 1000-3000 | open | |
| QET A/QET C | 3 | 7 | open | ||
| QET A/QET D | 3 | 38 | open | ||
| QET B/QET C | 35 | 7 | open | ||
| QET B/QET D | 35 | 38 | open | ||
| QET C/QET D | 7 | 38 | open | ||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | - | |
| LED 2 | chassis | 26 | - | ||
| Motor Coil X1 | 169 | OPEN | |||
| Motor Coil X2 | 169 | 0.038 k | |||
| Motor Coil Y1 | 169 | 0.033 k | |||
| Motor Coil Y2 | 169 | 0.020 k | |||
| Xendstop | 1k | - | |||
| Yendstop | 1k | 0.028 k | |||
| 1437 | Starting LHe transfer @MB=35% |
|---|---|
| 1525 | End LHe transfer @MB=94.8% |
| Leaving mixture in the fridge when starting 1K pot. Keep pumping on the 1Kpot during the next steps. | |
| 1552 | Sorb = 4.2 K. G1 (fridge) = 453 mbar. 1K Pot = 3.4 K / 0.5 mbar |
| 1555 | Open 1K Pot needle valve → 15° (half unit on the needle marks). 1K Pot → 2.96 K / 5.5 mbar. |
| 1556 | Open 1K Pot needle valve → 30°. 1K Pot → 2.25 K / 13 mbar. |
| 1600 | MC(H35) temp starts to decrease quickly, flattens at around 3 K |
| 1602 | MC(Cernox)/Still/MC(H35) = 3.21/2.99/3.14, 1K Pot = 2.34 K / 14 mbar. G1 (fridge+dump) = 333 mbar. |
| 1605 | Open 1K Pot needle valve → 45°. 1K Pot → 1.78 K / 36 mbar. |
| 1612 | MC(H35)/Still/1KPot = 2.65/2.48/1.67 K / 34 mbar, G1(fridge+dump) = 179 mbar |
| 1619 | Close V3 and V9, leaving 126 mbar in the dump. Turn on 3He pump. |
| 1622 | Start circulation through Trap 1. V6 → 12.4%. Condenser(G1) peaks at 287 mbar. P2 jumps up to 550 mbar, settling down to 60 mbar and then 28 mbar. |
| 1625 | V6 → 13.0%. Condenser(G1) didn't peak. MC(H35) → 1.06 K. Still(P1) → 1.11 K / 3 mbar. After some time the Condenser peaked even higher. Condenser(G1) peaks at 100 mbar |
|---|---|
| 1627 | V6 → 13.4%. Condenser(G1)didn't peak . MC(H35) → 0.99 K. Still(P1) → 1.11 K /1.4 mbar. P2 = 8 mbar |
| 1629 | V6 → 13.8%. Condenser(G1) didnt see a peak. MC(H35) → 0.89 K. Still(P1) → 1.03 K / 1.1 mbar. |
| 1631 | V6 → 14.4%. Condenser(G1) peaks at 83.5 mbar. MC(H35) → 0.82 K. Still(P1) → 1.01 K / 0.75 mbar. |
| 1632 | V6 → 15.0%. Condenser(G1) peaks at 87.5 mbar. MC(H35) → 0.76 K. Still(P1) → 1.02 K / 0.55 mbar. |
| 1634 | V6 → 100.%. Condenser(G1) peaks at 91.6 mbar. MC(H35) → 0.66 K. Still(P1) → 1.00 K / 0.22 mbar. |
| 1642 | Both G1 and P2 started increasing slowly. P2 went upto 60 mbar and settled back down to 30 mbar. G1 slowly creeping up |
| 1644 | MC(H35) = 208 mK, Still = 1.13 K /0.160 mbar, 1KPot = 1.64 K / 32 mbar, G1/G2 = 96/108 mbar |
| 1647 | Open V14 to let the remaining mix out of the dump and into circulation. P1(Still) went off-scale then falls again. MC(H35) peaks at 77 mK. Condenser peaks at 243 mbar. P2 peaks at 70 mbar. |
| 1701 | Tower(U28)/MC(H35)/Coldplate = 183/270/97 mK, Still(P1) = 1.08 K / 0.11 mbar / 0 mW, G1/G2 = 80/92 mbar, 1K Pot(P2) = 45° / 1.645 K / 30 mbar, MB = 94.2% |
| 1700 | Still (P1) < 4 mbar, close V14 |