start › cdms › k100 › run_diary › 20231031

2023/10/31 (Tuesday)

Previous Day

  • Lab in-personnel: Himangshu
Status Run 78 beginning. Started circulation yesterday. DCRCs off.
1700 Fill LN2 cold trap bath from 5L dewar
1716 Tower(U28)/MC(H35)/Coldplate = 75.9/29.8/145.9 mK, Still(P1) = 1.11 K / 0.11 mbar / 0 mW, G1/G2 = 128/116 mbar, 1K Pot(P2) = 37.5° / 1.65 K / 28 mbar, MB = 60.1%
1730 Check source mover motor coil resistances at base temperature : X1/X2/Y1/Y2 = 32.4/43.6/37/21.6 Ohm.
1542 Cable 2/1/3 → diverter/nothing/nothing → DCRC55(1)/DCRC50(3)/DCRCX(4)