UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20230213

2023/02/13 (Monday)

Previous Day

* Lab in-personnel: Ping, Zach, Himangshu

Status Run 77 in progress.
Setup To Be Cable 2/1/3 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4)
1121 Started refilling LN2
1133 Stopped refilling LN2
Cooling to base today
1310 Turn on 4He pump. Open Recovery valve to 4He pump. Pump out line to 1K pot (open V4a, close 1K pot speedivalve and slowly open the 4He-pump inlet valve).Close the 4He-pump inlet valve once done. Check pressure in 1K Pot (with V4a open, 4He pump inlet valve closed). 1K Pot = 0.23 bar on mechanical gauge.
1311 Pump out 1K Pot slowly, opening the 4He pump valve until it just starts to make noise, then opening it more when it becomes more quiet.
1316 Start pumping 1K Pot continuously
1338 Started LHe transfer @ MB = 33.9%
1344 MC(H35) = 4.46 K, cooling slowly. 1K Pot = 3.2 K / 1.5 mbar
1347 Turbo on IVC = 1 nbar, after being valved out for 1 min IVC = 10 mbar, close valves to IVC
1354 Open 1K Pot needle valve → 15° (half a unit mark). 1K Pot (P2) → 2.63 K / 8.0 mbar
1356 MC temperature falling MC(H35) = 3.57 K
1402 Open 1K Pot needle valve → 30°. 1K Pot → 1.85 K / 28 mbar. MC(H35) = 3.08 K
1413 MC(H35)/1Kpot = 2.69/1.66K still dropping linearly. G1/G2 = 193/202 mbar . 1K Pot(P2) = 26 mbar.
1428 P2 = 14 mbar now, Open 1K Pot needle valve → 45°. 1K Pot → 1.60 K / 18 mbar. MC(H35) = 2.14 K
1504 MC(H35)/1Kpot = 1.88/1.674 K still dropping linearly. G1/G2 = 49/52 mbar . 1K Pot(P2) = 40 mbar. jumped to 110 mbar and came back to 40 mbar and stable. Close back 1K Pot needle valve → 30°
1508 1K Pot(P2) = 30 mbar
1510 MC(H35)/1Kpot = 1.86/1.63 K still dropping linearly. G1/G2 = 48/51.8 mbar . 1K Pot(P2) = 28 mbar.
1511 Additional Step: Close V2 (only today). Close V3 and V9, leaving 51.9 mbar mixture in dump. P2 stable at 28 mbar.
1513 Turn on 3He pump
1514 Start circulation through Trap 1. V6 → 12.6%. Condenser peaks at 315 mbar. MC(H35) → 1.43 K. Still → 1.45 K / 14 mbar. Took about 2 minutes for P1 to start reading below max.
1528 V6 → 13.0%. Condenser(G1) flattened at 178.5 mbar. MC(H35) → 1.17 K. Still(P1) → 1.27 K / 5.0 mbar.
1531 Turned on all DCRCs.
1540 V6 → 13.4%. Condenser(G1) peaks at 189 mbar. MC(H35) → 1.08 K. Still(P1) → 1.18 K / 3.6 mbar.
1548 V6 → 13.8%. Condenser(G1) peaks at 192.6 mbar. MC(H35) → 0.88 K. Still(P1) → 1.19 K / 2.4 mbar.
1555 V6 → 14.4%. Condenser(G1) peaks at 194 mbar. MC(H35) → 0.60 K. Still(P1) → 1.16 K / 1.2 mbar.
1601 V6 → 15.0%. Condenser(G1) peaks at 187.7 mbar. MC(H35) → 0.33 K. Still(P1) → 1.16 K / 0.6 mbar.
1609 V6 → 100%. Condenser(G1) peaks at 184 mbar. MC(H35) → 0.145 K. Still(P1) → 1.12 K / 0.18 mbar. P2 remain steady at 28 mbar during the process of opening up V6
1654 MC(H35) = 31.6 mK and pretty flat, G1 = 125.0 mbar and pretty flat. 1K Pot = 1.64 K / 26 mbar.
1655 Open V14 to pull the rest of the mixture out of the dump into circulation.
1655 MC(H35) rises to 75 mK, Still peaks at 55 mbar then falls again, Condenser(G1) rises to 191 mbar. 1K Pot unaffected.
1659 MC(H35) rises further to 78 mK, then falls
To correct the mistake of opening the mixture to V2, Matt suggested opening V8 (which connects to the same volume) during this step
1703 P1 = 3.8 mbar, Open V8 to pump the mixture left in that line (from opening V2 before) into the fridge
1704 everything seems unaffected, P1 did not rise at all.
1709 MC temp rising again a small amount, currently at 42 mK from 36 mK
1721 P1 = 0.7 mbar (<1), close V8 and V14.
2036 Zapped DCRC4. I tried tuning the SQUIDS and found the following SQUIDS+channels working
DCRC1 PA/PB/PC/PD = (SQUID is not showing modulation)Side2pC/Side2pF/Side1pC/Side1pF
DCRC3 PA/PB/PC/PD = Side1pE /Side1pB /Side1pD/ Side2pE
DCRC4 PA/PB/PC/PD = (SQUID is not showing modulation)Side2pB/ (SQUID is not showing modulation)Side2pA/ Side1pA/ Side2pD
2121 Started miniIbIs points
From IbIs with MC = 30 mK
DCRC channel Detector channel QET bias @ dP/dI = - uV Previous Run
DCRC1pC Side1pC 235
DCRC1pD Side1pF 154 274
DCRC3pA Side1pE 140 246
DCRC3pB Side1pB 210 225
DCRC3pC Side1pD 350(no good pulses) 258
DCRC4pC Side1pA 212
DCRC1pA Side2pC
DCRC1pB Side2pF
DCRC3pD Side2pE
DCRC4pA Side2pB
DCRC4pB Side2pA
DCRC4pD Side2pD

some nexus ibis sequencer scripts and python code from Zach

Next Day