| Status | Run 77 in progress. Warmed up yesterday, first 10K pumpout. MC(H35) temp settled to 5.47 K overnight. |
| Setup To Be | Cable 2/1/3 → diverter/nothing/nothing → DCRC50(3)/DCRC55(1)/DCRCX(4) |
| 1455 | Started filling LN2 from the dewar. Checked with Mike to get trained for LN2 refill and found that the dewar still has 20L left |
| 1500 | Stopped filling LN2 |
| 1518 | Plugged in DCRC50(3) to diverter board to Cable 2. |
| 1538 | Tried changing the QETbias, Squid parameters but the DCRC is not responding |
| 1539 | Plugged in DCRC55(1) to nothing to Cable 1. All channels showing noise, PD shows some pulses as well!. |
| 1618 | Tried tuning the SQUIDS on DCRC1, SquidA donot show modulation. Other SQUIDs are fine and tuned. |
| 1621 | Plugged in DCRCX(4) to nothing to Cable 3. None of the channels are responding. |
| 1736 | DCRC4 showing traces on PA, PB and PC. PD shows traces sometimes and then goes away |